MORPHOMETRIC ANALYSIS OF THE CORNEAL ENDOTHELIUM WITH 3 DIFFERENT SPECULAR MICROSCOPES

被引:12
作者
LANDESZ, M
KAMPS, A
SLART, R
SIERTSEMA, JV
VANRIJ, G
机构
[1] Department of Ophthalmology, University Hospital Groningen, Groningen, 9700 RB
关键词
CELL MORPHOMETRY; CONTACT SPECULAR MICROSCOPY; CORNEAL ENDOTHELIUM; NONCONTACT SPECULAR MICROSCOPY;
D O I
10.1007/BF01203290
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The morphometry of the central corneal endothelium of 10 eyes in 10 subjects was analyzed with three different specular microscopes. Computer-assisted analysis was performed with only two microscopes (Zeiss and Keeler Konan sp 3300) because the third microscope (Topcon sp 1000) could not be adapted to our computerized system. With this Topcon microscope a grid with standard densities was used to compare the images with, in addition, we also performed manual cell counting on the same Topcon images. The coefficient of variation of the cell analysis of three different images per cornea with the four methods varied between 3.4 and 4.7 percent. One-way analysis of variance showed a significant difference between the Zeiss and the other microscopes. So only the Keeler Konan and the Topcon microscopes could be used interchangeably, The computerized image analysis permitted also an evaluation of the hexagonality. The results of polygonality were not significantly different between the Zeiss and the Keeler Konan. For clinical purposes the Topcon specular microscope is more advantageous than the other two methods, since it is the most rapid way to record and analyze specular images. But for more precise measurements an image processing system is indispensable.
引用
收藏
页码:15 / 28
页数:14
相关论文
共 17 条
[1]  
[Anonymous], 1991, Ophthalmology, V98, P1464
[2]   SPECULAR MICROSCOPY OF HUMAN CORNEAL ENDOTHELIUM INVIVO [J].
BOURNE, WM ;
KAUFMAN, HE .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1976, 81 (03) :319-323
[3]   MORPHOMETRY OF THE CORNEAL ENDOTHELIUM BY AUTOMATED TELEVISION IMAGE-ANALYSIS [J].
FABIAN, E ;
MERTZ, M ;
KODITZ, W .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1983, 182 (03) :218-223
[4]  
Hartmann C, 1984, Cornea, V3, P155
[5]  
Hirst L W, 1984, Cornea, V3, P83
[6]  
LAING RA, 1977, ARCH OPHTHALMOL-CHIC, V93, P143
[7]  
LESTER JM, 1981, INVEST OPHTH VIS SCI, V20, P407
[8]   CELLULAR MEMBRANE ACTIVITY IN CORNEAL ENDOTHELIUM OF INTACT EYE [J].
MAURICE, DM .
EXPERIENTIA, 1968, 24 (11) :1094-&
[9]  
NISHI O, 1989, CORNEA, V8, P54
[10]   AUTOMATED MORPHOMETRY OF CORNEAL ENDOTHELIAL-CELL - USE OF VIDEO CAMERA AND VIDEO TAPE-RECORDER [J].
NISHI, O ;
HANASAKI, K .