DIAGNOSIS OF PULMONARY-EDEMA BY A SURGICALLY NON-INVASIVE MICROWAVE TECHNIQUE

被引:23
作者
ISKANDER, MF [1 ]
DURNEY, CH [1 ]
SHOFF, DJ [1 ]
BRAGG, DG [1 ]
机构
[1] UNIV UTAH,DEPT RADIOL,SALT LAKE CITY,UT 84112
关键词
D O I
10.1029/RS014i06Sp00265
中图分类号
P1 [天文学];
学科分类号
0704 ;
摘要
A method for measuring content of water in the lung is described. The method utilizes changes in the amplitude and phase of the microwave signal transmitted through the thorax. The phase of the transmission coefficient is particularly sensitive and provides a means of detecting and monitoring of pulmonary edema. Theoretical predictions of sensitivity were confirmed experimentally in agar models of the thorax. The optimal frequency lies between 740 MHz and 1.5 GHz. A 50‐ohm open‐transmission line was designed and used to couple 915‐MHz energy into tissues with minimal external radiation. Other advantages of this strip‐line concept include its low mass, small size, and flat configuration, which render it utile for clinical purposes. The feasibility of the method was also evaluated in experiments on dogs. In a typical dog experiment, the phase of the microwave‐transmitted signal changed by an average value of 2° for each 10 milliliters of infused blood. These results provide strong encouragement that incipient pulmonary edema can be detected before clinical emergencies arise, thus providing the physician a safe and sensitive tool for diagnosis of pulmonary function. Copyright 1979 by the American Geophysical Union.
引用
收藏
页码:265 / 269
页数:5
相关论文
共 8 条
[1]   REBREATHING PULMONARY CAPILLARY AND TISSUE VOLUME IN NORMALS AFTER SALINE INFUSION [J].
FARNEY, RJ ;
MORRIS, AH ;
GARDNER, RM ;
ARMSTRONG, JD .
JOURNAL OF APPLIED PHYSIOLOGY, 1977, 43 (02) :246-253
[2]  
ISKANDER MF, 1977, J MICROWAVE POWER, V12, P16
[3]  
ISKANDER MF, 1979, P IEEE LETT SEP
[4]   PULMONARY EXTRAVASCULAR AND INTRAVASCULAR FLUID VOLUMES IN RESTING PATIENTS [J].
LUEPKER, R ;
LIANDER, B ;
KORSGREN, M ;
VARNAUSKAS, E .
AMERICAN JOURNAL OF CARDIOLOGY, 1971, 28 (03) :295-+
[5]   INVESTIGATION OF USE OF MICROWAVE RADIATION FOR PULMONARY DIAGNOSTICS [J].
PEDERSEN, PC ;
JOHNSON, CC ;
DURNEY, CH ;
BRAGG, DG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1976, 23 (05) :410-412
[6]   MICROWAVE REFLECTION AND TRANSMISSION MEASUREMENTS FOR PULMONARY DIAGNOSIS AND MONITORING [J].
PEDERSEN, PC ;
JOHNSON, CC ;
DURNEY, CH ;
BRAGG, DG .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1978, 25 (01) :40-48
[7]  
STAUB NC, 1974, PHYSIOL REV, V54, P678, DOI 10.1152/physrev.1974.54.3.678
[8]   POSSIBLE USE OF MICROWAVES IN MANAGEMENT OF LUNG-DISEASE [J].
SUSSKIND, C .
PROCEEDINGS OF THE IEEE, 1973, 61 (05) :673-674