MECHANISM OF BALLOON ANGIOPLASTY IN CHILDREN WITH ARTERIAL-STENOSIS ASSESSED BY INTRAVASCULAR ULTRASOUND AND ANGIOGRAPHY

被引:13
作者
INO, T [1 ]
OKUBO, M [1 ]
AKIMOTO, K [1 ]
NISHIMOTO, K [1 ]
YABUTA, K [1 ]
WATANABE, M [1 ]
HOSODA, Y [1 ]
机构
[1] JUNTENDO UNIV,SCH MED,DEPT THORAC SURG,BUNKYO KU,TOKYO 113,JAPAN
关键词
D O I
10.1016/0002-8703(95)90053-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Fifteen patients were examined before and immediately after balloon angioplasty with a 4.3F, 30 MHz rotational tip intravascular ultrasound system. In 12(80%) patients, 13 procedures could be analyzed because of sufficient image quality. The lesions for intravascular ultrasound (IVUS) study consisted of pulmonary arterial stenosis in 4 patients, Blalock-Taussig shunt stenosis in 3, coarctation of the aorta in 2, subclavian artery stenosis in 1, renal artery stenosis in 1, and ductus arteriosus in 1. Three distinctive morphologic types were identified: Group I had arterial stretching (3 patients); group Ita had superficial tearing (3); group IIb had deep intimal-medial tearing (5); group III had dissection (2). In the narrowest site, minimal and maximal luminal diameters, luminal area before and after balloon angioplasty were 3.5 +/- 1.8 mm vs 4.5 +/- 2.5 mm, 4.1 +/- 2.1 mm vs 5.4 +/- 3.5 mm, and 49.8 +/- 30.2 mm vs 88.3 +/- 45.2 mm(2), respectively. The recoil value of group IIb with appropriate balloon angioplasty was approximately 0.3. IVUS may be an useful modality for evaluating the morphologic mechanism of dilatation after balloon angioplasty.
引用
收藏
页码:132 / 138
页数:7
相关论文
共 9 条
[1]   TRANS-LUMINAL DILATATION OF COARCTATION OF THE ABDOMINAL-AORTA - AN EXPERIMENTAL-STUDY IN DOGS [J].
CASTANEDAZUNIGA, WR ;
LOCK, JE ;
VLODAVER, Z ;
RUSNAK, B ;
RYSAVY, JP ;
HERRERA, M ;
AMPLATZ, K .
RADIOLOGY, 1982, 143 (03) :693-697
[2]   CLASSIFICATION OF MORPHOLOGIC EFFECTS OF PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY ASSESSED BY INTRAVASCULAR ULTRASOUND [J].
GERBER, TC ;
ERBEL, R ;
GORGE, G ;
GE, J ;
RUPPRECHT, HJ ;
MEYER, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (20) :1546-1554
[3]   BALLOON ANGIOPLASTY OF COARCTATION OF THE AORTA EVALUATED WITH INTRAVASCULAR ULTRASOUND IMAGING [J].
HARRISON, JK ;
SHEIKH, KH ;
DAVIDSON, CJ ;
KISSLO, KB ;
LEITHE, ME ;
HIMMELSTEIN, SI ;
KANTER, RJ ;
BASHORE, TM .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) :906-909
[4]  
INO T, 1993, CARDIOL YOUNG, V3, P28
[5]  
JAIN A, 1991, AM HEART J, V123, P514
[6]   BALLOON DILATATION OF EXCISED AORTIC COARCTATIONS [J].
LOCK, JE ;
CASTANEDAZUNIGA, WR ;
BASS, JL ;
FOKER, JE ;
AMPLATZ, K ;
ANDERSON, RW .
RADIOLOGY, 1982, 143 (03) :689-691
[7]   HOW DOES ANGIOPLASTY WORK - SERIAL ANALYSIS OF HUMAN ILIAC ARTERIES USING INTRAVASCULAR ULTRASOUND [J].
LOSORDO, DW ;
ROSENFIELD, K ;
PIECZEK, A ;
BAKER, K ;
HARDING, M ;
ISNER, JM .
CIRCULATION, 1992, 86 (06) :1845-1858
[8]   INTRALUMINAL ULTRASOUND IMAGING THROUGH A BALLOON DILATION CATHETER IN AN ANIMAL-MODEL OF COARCTATION OF THE AORTA [J].
ROTHMAN, A ;
RICOU, F ;
WEINTRAUB, RG ;
CROWLEY, R ;
ARCANGELI, R ;
SHIOTA, T ;
ZHANG, J ;
SAHN, DJ .
CIRCULATION, 1992, 85 (06) :2291-2295
[9]   EVALUATION OF BALLOON ANGIOPLASTY OF COARCTATION RESTENOSIS BY MAGNETIC-RESONANCE-IMAGING [J].
SOULEN, RL ;
KAN, J ;
MITCHELL, S ;
WHITE, RI .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 60 (04) :343-345