CROSS-AXIS SYNCHRONOUS FLOW-THROUGH COIL PLANET CENTRIFUGE (TYPE-XLL) .2. SPECULATION ON THE HYDRODYNAMIC MECHANISM IN STATIONARY PHASE RETENTION

被引:10
作者
ITO, Y
机构
[1] Laboratory of Biophysical Chemistry, The National Heart, Lung, Blood Institute, Bethesda, MD 20892, Building 10
来源
JOURNAL OF CHROMATOGRAPHY | 1991年 / 538卷 / 01期
关键词
D O I
10.1016/S0021-9673(01)91622-X
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The hydrodynamic mechanism involved in the retention of stationary phase in the present x-axis coil planet centrifuge system is discussed. A statistical treatment of the retention data disclosed important clues such as the effect of the inward-outward elution mode, the close correlation between the planetary motion and the head-tail elution mode, and the superior retention capacity of the left-handed coils. The combined effects of the radial and lateral centrifugal force field derived from the mathematical analysis of acceleration acting on the coil provide an explanation for the phenomena.
引用
收藏
页码:67 / 80
页数:14
相关论文
共 8 条
[1]   HIGH-SPEED COUNTERCURRENT CHROMATOGRAPHY [J].
ITO, Y .
CRC CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1986, 17 (01) :65-143
[2]   CROSS-AXIS SYNCHRONOUS FLOW-THROUGH COIL PLANET CENTRIFUGE (TYPE-XLL) .1. DESIGN OF THE APPARATUS AND STUDIES ON RETENTION OF STATIONARY PHASE [J].
ITO, Y ;
KITAZUME, E ;
BHATNAGAR, M ;
TRIMBLE, FD .
JOURNAL OF CHROMATOGRAPHY, 1991, 538 (01) :59-66
[4]   CROSS-AXIS SYNCHRONOUS FLOW-THROUGH COIL PLANET CENTRIFUGE FREE OF ROTARY SEALS FOR PREPARATIVE COUNTERCURRENT CHROMATOGRAPHY .1. APPARATUS AND ANALYSIS OF ACCELERATION [J].
ITO, Y .
SEPARATION SCIENCE AND TECHNOLOGY, 1987, 22 (8-10) :1971-1987
[6]   CROSS-AXIS SYNCHRONOUS FLOW-THROUGH COIL PLANET CENTRIFUGE FOR LARGE-SCALE PREPARATIVE COUNTERCURRENT CHROMATOGRAPHY .1. APPARATUS AND STUDIES ON STATIONARY PHASE RETENTION IN SHORT COILS [J].
ITO, Y ;
ZHANG, TY .
JOURNAL OF CHROMATOGRAPHY, 1988, 449 (01) :135-151
[7]   EXPERIMENTAL-OBSERVATIONS OF THE HYDRODYNAMIC BEHAVIOR OF SOLVENT SYSTEMS IN HIGH-SPEED COUNTERCURRENT CHROMATOGRAPHY .3. EFFECTS OF PHYSICAL-PROPERTIES OF THE SOLVENT SYSTEMS AND OPERATING TEMPERATURE ON THE DISTRIBUTION OF 2-PHASE SOLVENT SYSTEMS [J].
ITO, Y ;
CONWAY, WD .
JOURNAL OF CHROMATOGRAPHY, 1984, 301 (02) :405-414
[8]  
ITO Y, 1988, COUNTERCURRENT CHROM, P79