ABSOLUTE LIGHT YIELD MEASUREMENTS ON BAF2 CRYSTALS AND THE QUANTUM EFFICIENCY OF SEVERAL PHOTOMULTIPLIER TUBES

被引:78
作者
DORENBOS, P
DEHAAS, JTM
VISSER, R
VANEIJK, CWE
HOLLANDER, RW
机构
[1] Radiation Technology Group, Department of Applied Physics, Delft University of Technology, 2629, JB Delft
关键词
D O I
10.1109/23.256593
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The quantum efficiency (QE) curves of five Philips XP2020Q photomultiplier tubes (PMTs), one Hamamatsu R2059 PMT, and a Na-Salicylate coated glass window were determined using a calibrated Thorn EMI 9426 PMT as a reference. The QE of XP2020Q PMTs at wavelengths smaller than 230 nm appears much better than the values specified by the manufacturer. Consequently, the often reported photon yield of 2000 photons/MeV for the fast component of the scintillation pulses from pure BaF2 crystals determined with this tube is overestimated. Our results, obtained by means of calibrated equipment and using 662 keV gamma-rays, yield a value of 1400 photons/MeV.
引用
收藏
页码:424 / 430
页数:7
相关论文
共 18 条
[1]   MEASUREMENT OF TIME DEPENDENCE OF SCINTILLATION INTENSITY BY A DELAYED-COINCIDENCE METHOD [J].
BOLLINGER, L ;
THOMAS, GE .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1961, 32 (09) :1044-+
[2]   BAF2 SCINTILLATORS WITH WIRE CHAMBER READOUT FOR POSITRON EMISSION TOMOGRAPHY [J].
BRUYNDONCKX, P ;
DEBRUYNE, J ;
ETIENNE, L ;
GRUWE, M ;
GUERARD, B ;
TAVERNIER, S ;
SHUPING, Z .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 310 (1-2) :107-115
[3]  
CHANG TB, 1987, NUCL INSTRUM METH A, V256, P398, DOI 10.1016/0168-9002(87)90239-7
[4]  
ERSHOV NN, 1982, OPT SPEKTROSK+, V53, P89
[5]   A MEASUREMENT OF THE LIGHT YIELD OF COMMON INORGANIC SCINTILLATORS [J].
HOLL, I ;
LORENZ, E ;
MAGERAS, G .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :105-109
[6]   LUMINESCENCE DUE TO RADIATIVE TRANSITIONS BETWEEN VALENCE BAND AND UPPER CORE BAND IN IONIC-CRYSTALS (CROSSLUMINESCENCE) [J].
JANSONS, JL ;
KRUMINS, VJ ;
RACHKO, ZA ;
VALBIS, JA .
PHYSICA STATUS SOLIDI B-BASIC RESEARCH, 1987, 144 (02) :835-844
[7]   PROPERTIES OF OPTICAL GREASES FOR BAF2 SCINTILLATORS [J].
KLAMRA, W ;
LINDBLAD, T ;
MOSZYNSKI, M ;
NORLIN, LO .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 254 (01) :85-87
[8]   BARIUM FLUORIDE INORGANIC SCINTILLATOR FOR SUBNANOSECOND TIMING [J].
LAVAL, M ;
MOSZYNSKI, M ;
ALLEMAND, R ;
CORMORECHE, E ;
GUINET, P ;
ODRU, R ;
VACHER, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 206 (1-2) :169-176
[9]   FURTHER STUDY OF SCINTILLATION-COUNTERS WITH BAF2 CRYSTALS FOR TIME-OF-FLIGHT POSITRON TOMOGRAPHY IN MEDICINE [J].
MOSZYNSKI, M ;
ALLEMAND, R ;
CORMORECHE, E ;
LAVAL, M ;
ODRU, R ;
VACHER, J .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1984, 226 (2-3) :534-541
[10]   RECENT MEASUREMENTS ON SCINTILLATOR-PHOTODETECTOR SYSTEMS [J].
SAKAI, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :418-422