REDUCTION OF FRETTING CORROSION OF TI-6AL-4V BY VARIOUS SURFACE TREATMENTS

被引:34
作者
MAURER, AM
BROWN, SA
PAYER, JH
MERRITT, K
KAWALEC, JS
机构
[1] CASE WESTERN RESERVE UNIV,DEPT BIOMED ENGN,10900 EUCLID AVE,CLEVELAND,OH 44106
[2] CASE WESTERN RESERVE UNIV,DEPT MAT SCI & ENGN,CLEVELAND,OH 44106
关键词
D O I
10.1002/jor.1100110613
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Titanium and titanium-6% aluminum-4% vanadium (Ti-6Al-4V) are known to-be biocompatible and corrosion resistant. However, there have been numerous reports of elevated tissue levels of titanium due to passive dissolution, wear, or fretting corrosion of implants. Studies were undertaken to determine whether the fretting corrosion of Ti-6Al-4V could be reduced by surface treatment of one or both surfaces in a fretting situation. Three different surface treatments were studied: ion implantation, physical vapor deposition nitriding, and plasma ion nitriding. The specimens used were screws fretting against the countersinks of a two-hole plate. Fretting corrosion was assessed by weight loss, by chemical analysis of test solutions, and by scanning electron microscopy. Surface treatment of one component, the screws, resulted in reduction in the release of titanium to only 18-32% of that seen with the untreated controls. Weight loss of the untreated plates fretted against physical vapor deposition nitrided screws and plasma ion nitrided screws was reduced to 31 and 38% of the control, respectively. The weight loss of plasma nitrided screws was only 30% that of the control. Nitriding of both plates and screws resulted in a further decrease in plate weight loss and metal release. Plasma ion nitriding of both components had the most significant effect, with the weight loss and titanium release being only 11 and 2% of the control values, respectively.
引用
收藏
页码:865 / 873
页数:9
相关论文
共 33 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]   METALLOSIS ASSOCIATED WITH A STABLE TITANIUM-ALLOY FEMORAL COMPONENT IN TOTAL HIP-REPLACEMENT - A CASE-REPORT [J].
BLACK, J ;
SHERK, H ;
BONINI, J ;
ROSTOKER, WR ;
SCHAJOWICZ, F ;
GALANTE, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (01) :126-130
[3]   FRETTING CORROSION IN SALINE AND SERUM [J].
BROWN, SA ;
MERRITT, K .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1981, 15 (04) :479-488
[4]   INVITRO STUDIES OF FRETTING CORROSION OF ORTHOPEDIC MATERIALS [J].
BROWN, SA ;
HUGHES, PJ ;
MERRITT, K .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1988, 6 (04) :572-579
[5]   CREVICE AND FRETTING CORROSION OF STAINLESS-STEEL PLATES AND SCREWS [J].
BROWN, SA ;
SIMPSON, JP .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1981, 15 (06) :867-878
[6]  
BROWN SA, 1985, CORROSION DEGRADATIO, P105
[7]   ION-IMPLANTATION OF SURGICAL TI-6AL-4V FOR IMPROVED RESISTANCE TO WEAR-ACCELERATED CORROSION [J].
BUCHANAN, RA ;
RIGNEY, ED ;
WILLIAMS, JM .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1987, 21 (03) :355-366
[8]   TRIBOLOGICAL PROPERTIES OF TITANIUM-ALLOYS [J].
BUDINSKI, KG .
WEAR, 1991, 151 (02) :203-217
[9]  
BUDINSKI KG, 1982, ASTM STP, V780, P49
[10]  
COLL B, 1990, 2ND INT ION NITRIDIN, P195