THE RELIABILITY OF AUSCULTATORY MEASUREMENT OF ARTERIAL BLOOD-PRESSURE - A COMPARISON OF THE STANDARD AND A NEW METHODOLOGY

被引:27
作者
STOLT, M
SJONELL, G
ASTROM, H
HANSSON, L
机构
[1] The Department of Development of Primary Health Care, Stockholm, Stockholm County Council West
[2] Kvarters-akuten, Stockholm
[3] Department of Clinical Physiology, Thoracic Clinics, Karolinska Hospital, Stockholm
[4] Department of Medicine, University of Göteborg, Östra Hospital, Göteborg
关键词
Auscultatory method; Blood pressure measurment; Tricuff;
D O I
10.1093/ajh/3.9.697
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
In 48 individuals with a wide range of arm circumferences blood pressure measured indirectly with two different cuffs was compared to direct intraarterial measurements. The two cuffs were a standard size cuff (12 X 35 cm) and a newly developed cuff, containing three rubber bags of different sizes, which automatically selects the appropriately sized bag in relation to arm circumference (Tricuff, Pressure Group AB, Sweden). The Tricuff correctly placed 42 of the 43 patients in the “normotensive” range, ie, diastolic blood pressure < 90 mm Hg, whereas the standard cuff put only 33 of 44 patients in this range (P <. 005). The discrepancy was not only due to the expected better performance in patients with obese arms, but also in the subgroup of patients with arms in the range 22 to 31 cm, in which group both cuffs would measure blood pressure with a 12 cm wide rubber bag. The number of correctly identified “normotensive” patients was noticeably higher with the Tricuff than with the standard cuff (30/31 v 25/32, P =. 053). The better specificity of the new cuff thus offers an improvement over the standard cuff. There are several potential clinical advantages of this, mainly that the risk of erroneously labelling normotensive individuals as hypertensive is reduced. © 1990 by the American Journal of Hypertension, Ltd.
引用
收藏
页码:697 / 703
页数:7
相关论文
共 18 条
[1]  
Frohlich E.D., Grim C., Labarthe D.R., Et al., Recommendations for human blood pressure determination by sphygmomanometers. Report of a special task force appointed by the steering committee, American Heart Association, Hypertension, 11, 2, pp. 210A-222A, (1988)
[2]  
Memorandum from a WHO/ISH meeting, Lancet, 2, pp. 457-458, (1983)
[3]  
Memorandumfrom a WHO/ISH meeting, J Hypertens, 4, pp. 383-386, (1986)
[4]  
Burke M.J., Towers H.M., O'Malley K., Et al., Sphygmomanometers in hospital and family practice: problems and recommendations, Brit Med J, 285, pp. 469-471, (1982)
[5]  
Maxwell M.H., Waks A.U., Schroth P.C., Et al., Error in blood pressure measurement due to incorrect cuff size in obese patients, Lancet, 2, pp. 33-36, (1982)
[6]  
Pickering G.W., High Blood Pressure, Churchill, (1968)
[7]  
Thulin T., Andersson G., Schersten B., Measurement of blood pressure—a routine test in need of standardization, Postgrad Med J, 51, pp. 390-395, (1975)
[8]  
Logan A.G., Report of the Canadian Hypertension Society‘s consensus conference on the management of mild hypertension, Can Med Assoc J, 131, pp. 1053-1057, (1984)
[9]  
Marking and rubber bag dimensions, (1970)
[10]  
Mancia G., Bertinieri G., Grassi G., Et al., Effects of blood pressure measurements by the doctor on patients blood pressure and heart rate, Lancet, i, pp. 695-698, (1983)