AN EVALUATION OF 3 LOADING CONFIGURATIONS FOR THE INVITRO TESTING OF FEMORAL STRAINS IN TOTAL HIP-ARTHROPLASTY

被引:37
作者
FINLAY, JB [1 ]
CHESS, DG [1 ]
HARDIE, WR [1 ]
RORABECK, CH [1 ]
BOURNE, RB [1 ]
机构
[1] DALHOUSIE UNIV,DEPT SURG,HALIFAX B3H 4H2,NS,CANADA
关键词
TOTAL HIP ARTHROPLASTY; FEMUR; STRAIN; LOADING CONFIGURATION; INVITRO;
D O I
10.1002/jor.1100090515
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Despite recent advances in total hip arthroplasty, proximal femoral resorption and osteopenia remain problems. To analyze the proximal strain effects of three different loading conditions, strains produced in intact and postarthroplasty femora have been compared. Ten adult cadaveric femora of similar size, shape, and rigidity were tested. Ten strain-gauge rosettes were positioned on each femur. To simulate the "single leg support" phase of gait, fixtures were developed to load the femora under three different configurations: the VS (vertical shaft) configuration, with a vertically orientated femur having rotational freedom proximally and distally; the ITB (iliotibial band) configuration of a femur with rotational freedom positioned 11-degrees from the vertical, with a strain-gauge adjustable metallic simulation of the lateral muscles of the thigh; and the ABD (abductor) configuration of a femur with rotational freedom positioned 11-degrees from the vertical, with a strain-gauge adjustable metallic simulation of the abductor muscles. Each femur was loaded less-than-or-equal-to 600 N through the medial point, located at one sixth of the transcondylar distance. Strain patterns and magnitudes produced by the three loading configurations were quite different in both the intact and postarthroplasty femora. Both the ITB and the ABD configurations resulted in greater proximal medial compression and lateral tension than did the VS configuration. The magnitudes of the proximal strains were significantly greater in the ABD configuration (p < 0.05). Postarthroplasty femora showed similar proximal results. It is proposed that meaningful strain data for the physiologically loaded femur can be obtained only with simulations that include the forces produced by the iliotibial band. To overcome the indeterminate nature and biological variation in these forces, the studies have to consider a range of forces.
引用
收藏
页码:749 / 759
页数:11
相关论文
共 31 条
[1]  
BOURNE RB, 1984, CLIN ORTHOP RELAT R, V183, P12
[2]   ANATOMY OF THE ABDUCTOR MUSCLES OF THE HIP AS STUDIED BY COMPUTED-TOMOGRAPHY [J].
CLARK, JM ;
HAYNOR, DR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1987, 69A (07) :1021-1031
[3]   BIOMECHANICAL INVESTIGATION OF HUMAN HIP [J].
CROWNINSHIELD, RD ;
JOHNSTON, RC ;
ANDREWS, JG ;
BRAND, RA .
JOURNAL OF BIOMECHANICS, 1978, 11 (1-2) :75-85
[4]   A MEASUREMENT OF PROXIMAL FEMUR STRAIN WITH TOTAL HIP-ARTHROPLASTY [J].
CROWNINSHIELD, RD ;
PEDERSEN, DR ;
BRAND, RA .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1980, 102 (03) :230-233
[5]   THE EFFECTS OF DRYING AND RE-WETTING ON SOME MECHANICAL-PROPERTIES OF CORTICAL BONE [J].
CURREY, JD .
JOURNAL OF BIOMECHANICS, 1988, 21 (05) :439-441
[6]  
DUNCAN JL, 1985, P JOINT C MATERIALS, P67
[7]   EFFECT OF FEMORAL COMPONENT SECTION MODULUS ON THE STRESS-DISTRIBUTION IN THE PROXIMAL HUMAN FEMUR [J].
ENGELHARDT, JA ;
SAHA, S .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1988, 26 (01) :38-45
[8]  
EVANS FG, 1948, ANAT REC, V100, P159
[9]   PROXIMAL FEMORAL SURFACE STRAIN-GAUGE ANALYSIS OF A NEW EPIPHYSEAL PROSTHESIS [J].
FIELD, RE ;
RUSHTON, N .
JOURNAL OF BIOMEDICAL ENGINEERING, 1989, 11 (02) :123-129
[10]  
Finlay J B, 1989, J Arthroplasty, V4, P335, DOI 10.1016/S0883-5403(89)80035-X