AVOIDING ATRIAL UNDERSENSING BY ASSESSMENT OF P-WAVE AMPLITUDE HISTOGRAM DATA

被引:19
作者
BOUTE, W
ALBERS, BA
GIELE, V
机构
[1] Vitatron Medical B.V, Velp
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1994年 / 17卷 / 11期
关键词
ATRIAL UNDERSENSING; DUAL CHAMBER PACEMAKER; HISTOGRAM;
D O I
10.1111/j.1540-8159.1994.tb03767.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Reliable sensing of the P wave is an essential requirement for the appropriate functioning of any device that uses atrial tracking to provide AV synchrony. However, a single measurement of the P wave amplitude, either at implantation or during follow-up, may not be a reliable reflection of the P wave amplitudes that occur during daily life. The P wave amplitude histogram is a new feature that automatically measures the P wave amplitude at regular intervals and provides the distribution of these measurements. This enables the assessment of the smallest P wave amplitudes that occur. Two populations were studied: 104 patients with a fixed atrial lead and a DDDR pacemaker and 100 patients with a single pass VDD lead and a VDD pacemaker. Both pacemakers incorporate the P wave amplitude histogram feature. Data in the P wave amplitude histogram were compared with a single measurement of the P wave amplitude at each follow-up. Programming of a 100% safety margin based on a single measurement of the P wave amplitude provided reliable atrial sensing in only 72% and 43% of the patients of both populations, respectively. Data continued in the P wave amplitude histogram may be a useful adjunct for the optimal programming of atrial sensitivity.
引用
收藏
页码:1878 / 1882
页数:5
相关论文
共 6 条
[1]   RELATIONSHIP OF QRS AMPLITUDE CHANGES DURING EXERCISE TO LEFT-VENTRICULAR FUNCTION AND VOLUMES AND THE DIAGNOSIS OF CORONARY-ARTERY DISEASE [J].
BATTLER, A ;
FROELICHER, V ;
SLUTSKY, R ;
ASHBURN, W .
CIRCULATION, 1979, 60 (05) :1004-1013
[2]   SENSOR-TRIGGERED, RATE-VARIABLE CARDIAC PACING - CURRENT TECHNOLOGIES AND CLINICAL IMPLICATIONS [J].
BENDITT, DG ;
MILSTEIN, S ;
BUETIKOFER, J ;
GORNICK, CC ;
MIANULLI, M ;
FETTER, J .
ANNALS OF INTERNAL MEDICINE, 1987, 107 (05) :714-724
[3]   AMPLITUDE AND DIRECTION OF ATRIAL DEPOLARIZATION USING A MULTIPOLAR FLOATING CATHETER - PRINCIPLES FOR A SINGLE LEAD VDD PACING [J].
FLAMMANG, D ;
RENIRIE, L ;
BEGEMANN, M ;
VANKRIEKEN, F .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1991, 14 (06) :1040-1048
[4]   ATRIAL SENSING PERFORMANCE OF AV UNIVERSAL PACEMAKERS DURING EXERCISE [J].
FROHLIG, G ;
BLANK, W ;
SCHWERDT, H ;
SEN, S ;
BETTE, L .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1988, 11 (01) :47-60
[5]  
LANGENFELD H, 1987, CARDIAC PACING ELECT, P185
[6]  
LEKVEN J, 1979, AM HEART J, V98, P200, DOI 10.1016/0002-8703(79)90222-9