摘要
A method is described in which a Farmer ionization chamber is used for the direct measurement of the air kerma rate in air from small spherical Caesium-137 sources at distances of 35 to 70 mm. The calibration factor and corrections for source and ion chamber geometry are examined. For quality assurance purposes, the results obtained are in sufficient agreement with other methods of calibration.
引用
收藏
页码:918 / 922
页数:5
相关论文
共 16 条
[1]   STUDIES OF CAVITY IONIZATION CHAMBERS WITH MEGAVOLTAGE X-RAYS [J].
BARNARD, GP ;
HITCHMAN, DG ;
MARSH, ARS .
PHYSICS IN MEDICINE AND BIOLOGY, 1964, 9 (03) :295-&
[2]  
EVANS RD, 1968, RAD DOSIMETRY, V1, P93
[3]  
EZZELL G, 1989, BRACHYTHERAPY, V2, P61
[4]   PHOTON MASS ATTENUATION AND ENERGY-ABSORPTION COEFFICIENTS FROM 1 KEV TO 20 MEV [J].
HUBBELL, JH .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1269-1290
[5]  
JONES CH, 1989, BRACHYTHERAPY 2, P75
[6]   EFFECT OF FINITE SIZE OF IONIZATION CHAMBERS ON MEASUREMENTS OF SMALL PHOTON SOURCES [J].
KONDO, S ;
RANDOLPH, ML .
RADIATION RESEARCH, 1960, 13 (01) :37-60
[7]  
LIQUORISH RAC, 1986, 54 54 I PHYS SCI MED
[8]   STANDARDIZATION OF CO-60 AND CS-137 GAMMA-RAY BEAMS IN TERMS OF EXPOSURE [J].
LOFTUS, TP ;
WEAVER, JT .
JOURNAL OF RESEARCH OF THE NATIONAL BUREAU OF STANDARDS SECTION A-PHYSICS AND CHEMISTRY, 1974, A 78 (04) :465-476
[9]  
MEERTENS H, 1984, 3RD P INT SEL US M I, P59
[10]  
READ LR, 1978, INT J APPL RADIAT IS, V29, P1