MAMMOGRAMS OBTAINED WITH RHODIUM VS MOLYBDENUM ANODES - CONTRAST AND DOSE DIFFERENCES

被引:16
作者
KIMMESMITH, C [1 ]
WANG, J [1 ]
DEBRUHL, N [1 ]
BASIC, M [1 ]
BASSETT, LW [1 ]
机构
[1] UNIV CALIF LOS ANGELES,SCH MED,BIOMED PHYS GRAD PROGRAM,LOS ANGELES,CA 90024
关键词
D O I
10.2214/ajr.162.6.8191989
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. A mammography unit with both a molybdenum anode and a rhodium anode, filtered with molybdenum and rhodium, respectively, was evaluated to determine which types of women would benefit from the dose savings of the rhodium combination despite some loss of contrast. SUBJECTS AND MATERIALS. In 100 women, the molybdenum anode and molybdenum filtration (Mo/Mo) were used to obtain mammograms of the right breast, and the rhodium anode and rhodium filtration (Rh/Rh) were used for mammograms of the left breast. All mammograms were obtained at 26 kVp. All milliampere-second values used to radiograph the breasts of these women were recorded. Mammograms of 54 women (30 with previous mammograms available), representing the four types of breasts as defined by the American College of Radiology, were interpreted by three radiologists. Each mammogram was assigned a grade for breast type, preference (Rh/Rh, Mo/Mo, or previous mammograms), contrast, and sharpness. RESULTS. Overall, mammograms obtained by using the Mo/Mo combination were preferred. However, for images of types 3 and 4 breasts, Rh/Rh was preferred twice as often as it had been for mammograms of types 1 and 2 breasts. The mean glandular dose for all breast types when the Rh/Rh combination was used was 42% of the dose used for the Mo/Mo combination. For a B-cm-thick dense breast, the Rh/Rh combination required 40% of the dose required for the Mo/Mo combination. CONCLUSION, Mammograms obtained with the Rh/Rh combination carried an overall decrease in contrast and mean glandular dose. However, for young women and some women with large dense breasts, the Rh/Rh mammograms were equivalent to or better than the mammograms obtained with the Mo/Mo combination. Effective use of Rh/Rh units requires careful selection of women based on age or the amount of glandular tissue seen on previous mammograms.
引用
收藏
页码:1313 / 1317
页数:5
相关论文
共 8 条
  • [1] BEAMAN SA, 1982, PHYS MED BIOL, V10, P1209
  • [2] INFLUENCE OF ANODE AND FILTER MATERIAL ON IMAGE QUALITY AND GLANDULAR DOSE FOR SCREEN FILM MAMMOGRAPHY
    DESPONDS, L
    DEPEURSINGE, C
    GRECESCU, M
    HESSLER, C
    SAMIRI, A
    VALLEY, JF
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (09) : 1165 - 1182
  • [3] OPTIMAL X-RAY-SPECTRA FOR SCREEN-FILM MAMMOGRAPHY
    JENNINGS, RJ
    EASTGATE, RJ
    SIEDBAND, MP
    ERGUN, DL
    [J]. MEDICAL PHYSICS, 1981, 8 (05) : 629 - 639
  • [4] X-RAY CHARACTERIZATION OF NORMAL AND NEOPLASTIC BREAST TISSUES
    JOHNS, PC
    YAFFE, MJ
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (06) : 675 - 695
  • [5] JOHNSON GA, 1976, RADIOLOGY, V127, P511
  • [6] SOME STATISTICAL-METHODS USEFUL IN CIRCULATION RESEARCH
    WALLENSTEIN, S
    ZUCKER, CL
    FLEISS, JL
    [J]. CIRCULATION RESEARCH, 1980, 47 (01) : 1 - 9
  • [7] WU X, 1993, MED PHYS, V20, P880
  • [8] SPECTRAL DEPENDENCE OF GLANDULAR TISSUE DOSE IN SCREEN-FILM MAMMOGRAPHY
    WU, XZ
    BARNES, GT
    TUCKER, DM
    [J]. RADIOLOGY, 1991, 179 (01) : 143 - 148