ATTENUATION OF ULTRASOUND IN SKELETAL-MUSCLE

被引:67
作者
NASSIRI, DK
NICHOLAS, D
HILL, CR
机构
[1] Physics Division, Institute of Cancer Research, Royal Marsden Hospital, Sutton, Surrey
关键词
Compendex;
D O I
10.1016/0041-624X(79)90054-4
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Ultrasonic attenuation in fresh and 5% formalin fixed beef skeletal muscle was measured, as a continuous function of frequency, in the range 1-8 MHz, for muscle fiber orientations both parallel and normal to the direction of propagation. Good agreement was found in all cases between 2 independent sets of measurements employing transmission and reflection techniques, respectively. A power law dependence of attenuation coefficient on frequency shows an exponent that is not significantly different from unity. For propagation normal to the fibers attenuation values are found as 1.1 .+-. 0.15 and 1.6 .+-. 0.15 dB cm-1 MHz-1 for fresh and fixed tissue, respectively, the corresponding values for parallel propagation being 2.9 .+-. 0.23 and 4.1 .+-. 0.25 dB cm-1 MHz-1.
引用
收藏
页码:230 / 232
页数:3