THE EFFECTS OF BRONCHODILATOR-INHALER AEROSOL PROPELLANTS ON RESPIRATORY GAS MONITORS

被引:8
作者
ELLIOTT, WR [1 ]
RAEMER, DB [1 ]
GOLDMAN, DB [1 ]
PHILIP, JH [1 ]
机构
[1] BRIGHAM & WOMENS HOSP,BOSTON,MA 02115
来源
JOURNAL OF CLINICAL MONITORING | 1991年 / 7卷 / 02期
关键词
MEASUREMENT TECHNIQUES; MASS SPECTROMETRY;
D O I
10.1007/BF01618119
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Spurious readings from a mass spectrometer have been reported following the administration of aerosol bronchodilators. We quantified the response of various respiratory gas analyzers to the aerosol propellant of albuterol inhalant (Proventil). The mass spectrometer systems tested, two Advantage systems, a SARA system, and a Model 6000 Ohmeda system, all displayed artifactual readings in response to the albuterol propellant. Each metered dose of the Proventil brand of albuterol contains 4 ml of Freon 11 (trichloromonofluoromethane) and 11 ml of Freon 12 (dichlorodifluoromethane). The concentration of propellant was expressed in doses/L, where each liter of gas contains 0.4 vol % of Freon 11 and 1.1 vol % of Freon 12 per dose. In proportion to the concentration of albuterol propellant, the two Advantage systems showed substantial readings of isoflurane (%) when no isoflurane was present (13% and 16% per dose/L) and reduced readings of enflurane (- 8% and - 10% per dose/L) and carbon dioxide (CO2) (- 3 and + 5 mm Hg per dose/L). The SARA system showed substantial CO2 readings when no CO2 was present (5 mm Hg per dose/L) and displayed small enflurane readings (0.1% per dose/L) when no enflurane was present. The Model 6000 unit showed CO2 readings when no CO2 was present (5 mm Hg per dose/L). Neither the Raman spectrometer, the infrared spectrometers, nor the piezoadsorptive analyzer we tested showed an artifactual effect of albuterol propellant on any of its readings. Simulation and clinical tests demonstrated that a single dose of albuterol propellant into a breathing circuit at the onset of inspiration resulted in concentrations of 0.8 and 0.3 dose/L, respectively. The phenomenon may be clinically useful, by allowing the anesthetist to verify the uptake of an inhalant into a patient.
引用
收藏
页码:175 / 180
页数:6
相关论文
共 4 条
[1]  
Gravenstein N., Theisen G.J., Knudsen A.K., Misleading mass spectrometer reading caused by an aerosol propellant, Anesthesiology, 62, pp. 70-72, (1985)
[2]  
McCleary U., Potential effects of an unknown gas on mass spectrometer readings, Anesthesiology, 63, pp. 724-725, (1985)
[3]  
Knudsen A.K., Johnson J.V., Yost R.A., More on mass spectrometers and aerosol propellants, Anesthesiology, 63, pp. 568-569, (1985)
[4]  
Severinghaus J.W., Young W.G., A gas mixer for computer calibration of anesthetic mass spectrometers, J Clin Monit, 2, pp. 223-229, (1986)