BYPASSING FEMORAL CORTICAL DEFECTS WITH CEMENTED INTRAMEDULLARY STEMS

被引:77
作者
LARSON, JE
CHAO, EYS
FITZGERALD, RH
机构
[1] MAYO CLIN & MAYO FDN,DEPT ORTHOPED,ROCHESTER,MN 55905
[2] MAYO CLIN & MAYO FDN,BIOMECH LAB,ROCHESTER,MN 55905
[3] WAYNE STATE UNIV,DEPT ORTHOPED SURG,DETROIT,MI 48202
关键词
CORTICAL DEFECT; FEMORAL PROSTHESIS; STEM BYPASS;
D O I
10.1002/jor.1100090314
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The objective of this investigation was to examine the effect of cemented intramedullary stem bypass on bone torsional property in the presence of femoral cortical defect. We intended to test two hypotheses; first, intramedullary fixation without bypass will accentuate the stress concentration effect and second, there will be an optimal length of stem bypass beyond the defect. Cemented intramedullary stems were used to bypass 50% diaphyseal diameter unicortical defects in paired, fresh-frozen canine femora. One member of each pair served as an unaltered control and all specimens were tested to failure in torsion. Both single-tailed paired t-test and analysis of variance were used for data analysis. Bones subjected to the cortical defect and no bypass were substantially weakened, exhibiting only 44 +/- 8% of control side maximum torque (p < 0.001). Positioning of the intramedullary stem with its tip at the center of the defect provided a small degree of strength improvement, achieving 60 +/- 7% of control side maximum torque. One and two diaphyseal diameter bypass, however, significantly (p < 0.01) improved bone torsional strength, resulting in 80 +/- 6% and 84 +/- 13% of control side maximum torque, respectively. Three diameter bypass achieved only 68 +/- 8% of control side maximum torque. This significant (p < 0.05) decline in strength when compared to two diameter bypass appears to indicate that the length of stem bypass beyond the cortical defect does have an optimum, probably due to the geometric characteristics of the canine femora.
引用
收藏
页码:414 / 421
页数:8
相关论文
共 30 条
[1]
AMSTUTZ HC, 1982, CLIN ORTHOP RELAT R, V170, P21
[2]
BECHTOL CO, 1952, AAOS INSTRUCTIONAL C
[3]
BETHEA JS, 1982, CLIN ORTHOP RELAT R, V170, P95
[4]
BIOMECHANICS OF TORSIONAL FRACTURES - STRESS CONCENTRATION EFFECT OF A DRILL HOLE [J].
BROOKS, DB ;
BURSTEIN, AH ;
FRANKEL, VH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1970, A 52 (03) :507-&
[5]
BONE STRENGTH - EFFECT OF SCREW HOLES [J].
BURSTEIN, AH ;
VESSELY, JC ;
FRANKEL, VH ;
LUNSETH, P ;
CURREY, J ;
HEIPLE, KG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1972, 54 (06) :1143-+
[6]
STANDARD TEST FOR LABORATORY ANIMAL BONE [J].
BURSTEIN, AH ;
FRANKEL, VH .
JOURNAL OF BIOMECHANICS, 1971, 4 (02) :155-&
[7]
RESULTS OF REVISION FOR MECHANICAL FAILURE AFTER CEMENTED TOTAL HIP-REPLACEMENT, 1979 TO 1982 - A 2 TO 5-YEAR FOLLOW-UP [J].
CALLAGHAN, JJ ;
SALVATI, EA ;
PELLICCI, PM ;
WILSON, PD ;
RANAWAT, CS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (07) :1074-1085
[8]
EFFECT OF BIOPSY-HOLE SHAPE AND SIZE ON BONE STRENGTH [J].
CLARK, CR ;
MORGAN, C ;
SONSTEGARD, DA ;
MATTHEWS, LS .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (02) :213-217
[9]
EDGERTON BC, 1989, THESIS MAYO GRADUATE
[10]
INCIDENCE OF FRACTURE THROUGH METASTASES IN LONG BONES [J].
FIDLER, M .
ACTA ORTHOPAEDICA SCANDINAVICA, 1981, 52 (06) :623-627