THE EFFECT OF ENERGY-DEPENDENT DETECTOR EFFICIENCY IN ELECTRON SPECTROSCOPIC METHODS - XPS, AES AND DCEMS

被引:4
作者
KAJCSOS, Z [1 ]
MEISEL, W [1 ]
GRIESBACH, P [1 ]
GUTLICH, P [1 ]
机构
[1] UNIV MAINZ,INST INORGAN & ANALYT CHEM,W-6500 MAINZ,GERMANY
关键词
D O I
10.1002/sia.740200610
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Systematic studies were performed to investigate the distorting effect of the electron energy dependence of the channeltron detector efficiency on the shape of data distributions in electron spectroscopy. In agreement with earlier results for electron energies up to 3 keV, it was found that a virtually energy-independent shape of the electron distribution might be achieved even in a very broad range of electron energies (up to 15 keV) by appropriate selection of the operational high voltage of a channeltron. However, the efficiency of channeltrons was also found to depend on count rates, indicating one of the principal difficulties in comparing distributions recorded under different experimental conditions.
引用
收藏
页码:544 / 548
页数:5
相关论文
共 19 条
[1]   METHOD OF ANALYSIS OF THIN SURFACE LAYERS BY MOSSBAUER EFFECT [J].
BONCHEV, Z ;
JORDANOV, A ;
MINKOVA, A .
NUCLEAR INSTRUMENTS & METHODS, 1969, 70 (01) :36-&
[2]  
FEDOSOV VV, 1972, THEORY OPTIMAL EXPT
[3]   IMPROVEMENTS IN FE-57 DEPTH-SELECTIVE CONVERSION ELECTRON MOSSBAUER-SPECTROSCOPY BY APPLYING A LOW-NOISE SCINTILLATION ELECTRON DETECTOR [J].
KAJCSOS, Z ;
SAUER, C ;
HOLZWARTH, A ;
KURZ, R ;
ZINN, W ;
LIGTENBERG, MAC ;
VANALLER, G .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1988, 34 (03) :383-390
[4]   ICEMS AND DCEMS STUDY OF FE LAYERS EVAPORATED ONTO AL AND SI [J].
KAJCSOS, Z ;
MEISEL, W ;
KUZMANN, E ;
TOSELLO, C ;
GRATTON, ML ;
VERTES, A ;
GUTLICH, P ;
NAGY, DL .
HYPERFINE INTERACTIONS, 1990, 57 (1-4) :1883-1888
[5]   PROBLEMS OF ELECTRON DETECTION IN DEPTH-SELECTIVE CONVERSION ELECTRON MOSSBAUER-SPECTROSCOPY [J].
KAJCSOS, Z ;
MEISEL, W ;
GRIESBACH, P ;
GUTLICH, P ;
LIGTENBERG, MAC ;
GRADMANN, U ;
LEHRBERGER, E ;
PRZYBYLSKI, M .
HYPERFINE INTERACTIONS, 1992, 71 (1-4) :1483-1486
[6]   SCINTILLATION DETECTORS IN CONVERSION ELECTRON MOSSBAUER-SPECTROSCOPY [J].
KAJCSOS, Z ;
SAUER, C ;
ZINN, W ;
LIGTENBERG, MAC .
HYPERFINE INTERACTIONS, 1992, 71 (1-4) :1479-1482
[7]   MEASUREMENT OF THE DETECTION EFFICIENCY OF MICROCHANNEL PLATES FOR 1-15 KEV ELECTRONS [J].
KLINGELHOEFER, G ;
WIACKER, H ;
KANKELEIT, E .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 247 (02) :379-384
[8]   STUDY OF VERY THIN OXIDE LAYERS BY CONVERSION AND AUGER ELECTRONS [J].
KLINGELHOFER, G ;
MEISEL, W .
HYPERFINE INTERACTIONS, 1990, 57 (1-4) :1911-1918
[9]  
LEE RG, 1957, PHYS REV, V107, P979
[10]   ANALYSIS OF ELECTRON-TRANSPORT IN CONVERSION ELECTRON MOSSBAUER-SPECTROSCOPY (CEMS) [J].
LILJEQUIST, D ;
EKDAHL, T ;
BAVERSTAM, U .
NUCLEAR INSTRUMENTS & METHODS, 1978, 155 (03) :529-538