MAGNETIC RESONANCE IMAGING;
MAGNETIC RESONANCE ANGIOGRAPHY;
BLOOD FLOW;
SLICE PROFILE;
D O I:
10.1002/mrm.1910290610
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
We introduce an asymmetric slice profile technique, which alters the spatial response of the flow signal in 3D time-of-flight NMR angiography. By gradually increasing the flip angle from the inflow to the outflow portions of the slab, the inflow refreshment effect is distributed over a wide slab thickness. The asymmetric slice profile is simply produced by using a Gaussian RF excitation with an overlapping presaturation. The spatial distribution of steady flow signal in a phantom study demonstrated an essential agreement with a numerical simulation. 3D time-of-flight NMR angiography of volunteers' heads using this technique provided a smooth vascular depiction over a wide slab thickness.