3-DIMENSIONAL REALIGNMENT OF ACTIVATION BRAIN SINGLE-PHOTON EMISSION TOMOGRAPHIC STUDIES

被引:18
作者
PAVIA, J [1 ]
ROS, D [1 ]
CATAFAU, AM [1 ]
LOMENA, FJ [1 ]
HUGUET, M [1 ]
SETOAIN, J [1 ]
机构
[1] UNIV BARCELONA,FAC MED,BIOPHYS & BIOENGN LAB,BARCELONA,SPAIN
来源
EUROPEAN JOURNAL OF NUCLEAR MEDICINE | 1994年 / 21卷 / 12期
关键词
REGISTERING; REALIGNMENT; ACTIVATION BRAIN STUDIES; CORRELATION; PRINCIPAL AXES TRANSFORMATION;
D O I
10.1007/BF02426693
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Two methods for 3D realignment of activation brain single-photon emission tomographic (SPET) studies are analyzed. The first is based on principal axes transformation (PAT). The second uses the results of the first method as initial values to start a least-squares iterative process (LS) to search for the maximum value of the correlation function. Both methods were tested with simulated and real studies. The results of the PAT method showed a maximum translation error of 0,3 +/- 0.1 pixels and a rotational error of 1.2 +/- 0.7 degrees in a total of 100 runs. For the LS method these errors were 0.2 +/- 0.1 and 0.6 +/- 0.3. The realignment for 34 real studies was assessed by three expert observers. The alignment was found to be satisfactory in all cases for the LS method, and in 18 cases (53%) for the PAT method. From the results we conclude that a combination of both methods allows the accurate realignment of SPET neuroactivation studies.
引用
收藏
页码:1298 / 1302
页数:5
相关论文
共 12 条
[1]  
ALPERT NM, 1990, J NUCL MED, V31, P1717
[2]  
CATAFAU AM, 1994, J NUCL MED, V35, P935
[3]  
GOLDSTEIN H, 1959, CLASSICAL MECHANICS
[4]  
JUNCK L, 1990, J NUCL MED, V31, P1220
[5]   EVALUATION OF THALLIUM MYOCARDIAL-PERFUSION IMAGING BY MEANS OF WASHOUT PARAMETRIC IMAGES - A SIMULATION STUDY [J].
JUVELLS, I ;
ROS, D ;
PAVIA, J ;
GARCIA, A .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1991, 8 (01) :157-163
[6]   RETROSPECTIVE GEOMETRIC CORRELATION OF MR, CT, AND PET IMAGES [J].
LEVIN, DN ;
PELIZZARI, CA ;
CHEN, GTY ;
CHEN, CT ;
COOPER, MD .
RADIOLOGY, 1988, 169 (03) :817-823
[7]  
MINOSHIMA S, 1992, J NUCL MED, V33, P1579
[8]   A SIMPLEX-METHOD FOR FUNCTION MINIMIZATION [J].
NELDER, JA ;
MEAD, R .
COMPUTER JOURNAL, 1965, 7 (04) :308-313
[9]   ACCURATE 3-DIMENSIONAL REGISTRATION OF CT, PET, AND OR MR IMAGES OF THE BRAIN [J].
PELIZZARI, CA ;
CHEN, GTY ;
SPELBRING, DR ;
WEICHSELBAUM, RR ;
CHEN, CT .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1989, 13 (01) :20-26
[10]   DETERMINATION OF THE RELATIVE ORIENTATION BETWEEN 2 IMAGES BY CORRELATION OF UNIDIMENSIONAL FUNCTIONS [J].
PLADELLORENS, J ;
JUVELLS, I ;
ROS, D .
JOURNAL OF OPTICS-NOUVELLE REVUE D OPTIQUE, 1987, 18 (04) :203-208