EFFECT OF STEM MODULUS IN A TOTAL HIP-ARTHROPLASTY MODEL

被引:52
作者
MAISTRELLI, GL
FORNASIER, V
BINNINGTON, A
MCKENZIE, K
SESSA, V
HARRINGTON, I
机构
[1] UNIV TORONTO,TORONTO E GEN & ORTHOPAED HOSP,TORONTO M4C 3E7,ONTARIO,CANADA
[2] UNIV GUELPH,ONTARIO VET COLL,GUELPH N19 2W1,ONTARIO,CANADA
[3] UNIV TORONTO,PRINCESS MARGARET HOSP,TORONTO M4X 1K9,ONTARIO,CANADA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1991年 / 73卷 / 01期
关键词
D O I
10.1302/0301-620X.73B1.1991773
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of this study was to determine the biological effects of the elastic modulus of the femoral stem in canine hip arthroplasty. Cementless total hip arthroplasty was performed in 12 dogs, six had a low elastic modulus polyacetal resin stem and six had a high modulus stainless steel stem. The components were otherwise similar. At six and 12 months after operation, radiographic and histomorphometric analysis showed that those with steel implants had more cortical porosity than did the other group (p < 0.01). We suggest that the elastic modulus of the implant is an important factor in controlling cortical bone resorption. A low modulus femoral prosthesis can significantly decrease bone resorption which might otherwise eventually lead to implant failure.
引用
收藏
页码:43 / 46
页数:4
相关论文
共 20 条
[1]   OSTEOLYTIC CHANGES IN THE UPPER FEMORAL-SHAFT FOLLOWING POROUS-COATED HIP-REPLACEMENT [J].
BROWN, IW ;
RING, PA .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1985, 67 (02) :218-221
[2]   THE EFFECTS OF INTRA-MEDULLARY IMPLANTS ON BONE STRAINS AND REMODELING IN THE FEMUR [J].
COOK, SD ;
WEINSTEIN, AM ;
LUEDEMANN, R ;
LAVERNIA, C ;
SKINNER, HB ;
KLAWITTER, JJ .
BIOMATERIALS MEDICAL DEVICES AND ARTIFICIAL ORGANS, 1982, 10 (01) :21-40
[3]  
Crowninshield R D, 1981, Hip, P111
[4]   CALCAR UNLOADING AFTER HIP-REPLACEMENT - A CADAVER STUDY OF FEMORAL STEM DESIGNS [J].
DJERF, K ;
GILLQUIST, J .
ACTA ORTHOPAEDICA SCANDINAVICA, 1987, 58 (02) :97-103
[5]   POROUS-COATED HIP-REPLACEMENT - THE FACTORS GOVERNING BONE INGROWTH, STRESS SHIELDING, AND CLINICAL-RESULTS [J].
ENGH, CA ;
BOBYN, JD ;
GLASSMAN, AH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1987, 69 (01) :45-55
[6]  
ENGH CA, 1988, CLIN ORTHOP RELAT R, V231, P7
[7]  
HUISKES R, 1984, FUNCTIONAL BEHAVIOR, V2, P121
[8]   INVESTIGATION INTO BIOMECHANICAL CAUSES OF PROSTHESIS STEM LOOSENING WITHIN THE PROXIMAL END OF THE HUMAN FEMUR [J].
JACOB, HAC ;
HUGGLER, AH .
JOURNAL OF BIOMECHANICS, 1980, 13 (02) :159-173
[9]  
KUSSWETTER H, 1984, CEMENTLESS FIXATION, P17
[10]  
LEWIS J, 1981, Orthopaedic Transactions, V5, P245