SPECTROPHOTOMETRIC STOPPED-FLOW APPARATUS SUITABLE FOR HIGH-PRESSURE EXPERIMENTS TO 200-MPA

被引:180
作者
VANELDIK, R [1 ]
GAEDE, W [1 ]
WIELAND, S [1 ]
KRAFT, J [1 ]
SPITZER, M [1 ]
PALMER, DA [1 ]
机构
[1] UNIV FRANKFURT,INST PHYS CHEM,W-6000 FRANKFURT,GERMANY
关键词
D O I
10.1063/1.1144466
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A modified high-pressure, stopped-flow apparatus has been developed to enable kinetic experiments to be monitored spectrophotometrically to 200 MPa, with efficient mixing of the reactant solutions. This apparatus is compact, relatively inexpensive to construct, and the solutions are only exposed to inert materials, such as Kel-F, Teflon, quartz (low-pressure windows), and glass or Viton. The drive-syringe pistons are propelled by a step motor housed on top of the stopped-flow unit inside the pressure vessel, resulting in a dead time for the system of approximately 10 ms at 25-degrees-C.