DETECTOR RESPONSE RESTORATION IN IMAGE-RECONSTRUCTION OF HIGH-RESOLUTION POSITRON EMISSION TOMOGRAPHY

被引:51
作者
LIANG, ZG
机构
[1] Department of Radiology, State University of New York, Stony Brook
关键词
D O I
10.1109/42.293924
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
A mathematical method was studied to model the detector response of high spatial-resolution positron emission tomography systems consisting of close-packed small crystals, and to restore the resolution deteriorated due to crystal penetration and/or nonuniform sampling across the field-of-view (FOV). The simulated detector system had 600 bismuth germanate crystals of 3.14 mm width and 30 mm length packed on a single ring of 60 cm diameter. The space between crystals was filled up with lead (i.e., septa). Each crystal was in coincidence with 200 opposite crystals so that the FOV had a radius of 30 cm. The detector response was modeled based on the attenuating properties of the crystals and the septa, as well as the geometry of the detector system. The modeled detector-response function was used to restore the projections from the sinogram of the ring-detector system. The restored projections had a uniform sampling of 1.57 mm across the FOV. The crystal penetration and/or the nonuniform sampling were compensated in the projections. A penalized maximum-likelihood algorithm was employed to accomplish the restoration. The restored projections were then filtered and backprojected to reconstruct the image. A chest phantom with a few small circular ''cold'' objects (almost-equal-to 4 mm diameter) located at the center and near the periphery of FOV was computer generated and used to test the restoration. The reconstructed images from the restored projections demonstrated resolution improvement off the FOV center, while preserving the resolution near the center.
引用
收藏
页码:314 / 321
页数:8
相关论文
共 34 条
[1]   A PET DETECTOR WITH DEPTH-OF-INTERACTION DETERMINATION [J].
BARTZAKOS, P ;
THOMPSON, CJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (06) :735-748
[2]   CORRECTION FOR SCATTERED RADIATION IN A RING DETECTOR POSITRON CAMERA BY INTEGRAL TRANSFORMATION OF THE PROJECTIONS [J].
BERGSTROM, M ;
ERIKSSON, L ;
BOHM, C ;
BLOMQVIST, G ;
LITTON, J .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1983, 7 (01) :42-50
[3]   EMISSION COMPUTER-ASSISTED TOMOGRAPHY WITH SINGLE-PHOTON AND POSITRON-ANNIHILATION PHOTON EMITTERS [J].
BUDINGER, TF ;
DERENZO, SE ;
GULLBERG, GT ;
GREENBERG, WL ;
HUESMAN, RH .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1977, 1 (01) :131-145
[4]   DESIGN OF A HIGH-RESOLUTION POSITRON EMISSION TOMOGRAPH USING SOLID-STATE SCINTILLATION DETECTORS [J].
CARRIER, C ;
MARTEL, C ;
SCHMITT, D ;
LECOMTE, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :685-690
[5]   HIGH-RESOLUTION POSITRON EMISSION TOMOGRAPHY USING SMALL BISMUTH-GERMANATE CRYSTALS AND INDIVIDUAL PHOTOSENSORS [J].
DERENZO, SE ;
BUDINGER, TF ;
VULETICH, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1983, 30 (01) :665-670
[6]   INITIAL CHARACTERIZATION OF A POSITION-SENSITIVE PHOTODIODE BGO DETECTOR FOR PET [J].
DERENZO, SE ;
MOSES, WW ;
JACKSON, HG ;
TURKO, BT ;
CAHOON, JL ;
GEYER, AB ;
VULETICH, T .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1989, 36 (01) :1084-1089
[7]   INITIAL RESULTS FROM THE DONNER-600 CRYSTAL POSITRON TOMOGRAPH [J].
DERENZO, SE ;
HUESMAN, RH ;
CAHOON, JL ;
GEYER, A ;
UBER, D ;
VULETICH, T ;
BUDINGER, TF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :321-325
[8]   MATHEMATICAL REMOVAL OF POSITRON RANGE BLURRING IN HIGH-RESOLUTION TOMOGRAPHY [J].
DERENZO, SE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1986, 33 (01) :565-569
[9]   A POSITRON TOMOGRAPH WITH 600-BGO CRYSTALS AND 2.6 MM RESOLUTION [J].
DERENZO, SE ;
HUESMAN, RH ;
CAHOON, JL ;
GEYER, AB ;
MOSES, WW ;
UBER, DC ;
VULETICH, T ;
BUDINGER, TF .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1988, 35 (01) :659-664
[10]   ANALYSIS OF POSITRON-EMISSION TOMOGRAPHY SCINTILLATION-DETECTORS WITH WEDGE FACES AND INTER-CRYSTAL SEPTA [J].
HOLMES, TJ ;
FICKE, DC .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1985, 32 (01) :826-830