DETERMINATION OF CARPROFEN IN BLOOD BY GAS-CHROMATOGRAPHY CHEMICAL IONIZATION MASS-SPECTROMETRY

被引:7
作者
HODSHON, BJ [1 ]
GARLAND, WA [1 ]
PERRY, CW [1 ]
BADER, GJ [1 ]
机构
[1] HOFFMANN LA ROCHE INC,DEPT CHEM RES,NUTLEY,NJ 07110
来源
BIOMEDICAL MASS SPECTROMETRY | 1979年 / 6卷 / 08期
关键词
D O I
10.1002/bms.1200060803
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A gas chromatographic chemical ionization mass spectrometric assay has been developed to measure carprofen in blood. The method features the addition of either a structural or stable isotope analog internal standard to plasma prior to a simple benzene extraction at pH 4.5. The residue, after removal of the benzene, is methylated with ethereal diazomethane. Following evaporation of the methylating solvents, a portion of the reconstituted residue is analyzed by gas chromatography mass spectrometry. The mass spectrometer is set to monitor in the gas chromatographic effluent the [MH]+ions of carprofen methyl ester and the methyl ester of internal standard generated by isobutane chemical ionization. Assay sensitivity is 5 pmol ml−1. When 200 pmol ml−1 samples are analyzed using a stable isotope analog as the internal standard, the precision and accuracy are both 4%. Using a structural analog as the internal standard, the assay was neither as precise nor as accurate. Copyright © 1979 Heyden & Son Ltd
引用
收藏
页码:325 / 330
页数:6
相关论文
共 21 条
[1]  
BEYNON JH, 1963, MASS ABUNDANCE TABLE, P118
[2]   DETERMINATION OF PLASMA TESTOSTERONE BY COMBINED GAS CHROMATOGRAPHY MASS SPECTROMETRY [J].
CHAPMAN, JR ;
BAILEY, E .
JOURNAL OF CHROMATOGRAPHY, 1974, 89 (02) :215-224
[3]   VARIANCE ANALYSIS OF ERROR IN SELECTED ION MONITORING ASSAYS USING VARIOUS INTERNAL STANDARDS - PRACTICAL STUDY CASE [J].
CLAEYS, M ;
MARKEY, SP ;
MAENHAUT, W .
BIOMEDICAL MASS SPECTROMETRY, 1977, 4 (02) :122-128
[4]  
DUTCHER JS, 1977, CLIN PHARMACOL THER, V22, P447
[5]   CHEMICAL IONIZATION MASS SPECTROMETRY .9. TEMPERATURE AND PRESSURE STUDIES WITH BENZYL ACETATE AND T-AMYL ACETATE [J].
FIELD, FH .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 1969, 91 (11) :2827-+
[6]  
GARLAND WA, 1979, CLIN PHARMACOL THER, V25, P844
[7]   DETERMINATION OF IMIPRAMINE IN PLASMA BY HIGH-PRESSURE LIQUID-CHROMATOGRAPHY AND FIELD-IONIZATION MASS-SPECTROMETRY - INCREASED SENSITIVITY IN COMPARISON WITH GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
HECK, HA ;
FLYNN, NW ;
BUTTRILL, SE ;
DYER, RL ;
ANBAR, M .
BIOMEDICAL MASS SPECTROMETRY, 1978, 5 (03) :250-257
[8]  
HOLLAND JF, 1973, 21ST ANN C MASS SPEC, P55
[9]   COMPARISON OF UNLABELLED AND LABELED INTERNAL STANDARDS FOR QUANTIFICATION BY SINGLE AND MULTIPLE ION MONITORING [J].
LEE, MG ;
MILLARD, BJ .
BIOMEDICAL MASS SPECTROMETRY, 1975, 2 (02) :78-81
[10]  
METZLER CM, USERS MANUAL NONLIN