A NEW MODEL TO TEST PLATELET-ADHESION UNDER DYNAMIC CONDITIONS - APPLICATION TO THE EVALUATION OF A TITANIUM NITRIDE COATING

被引:21
作者
DION, I
BAQUEY, C
HAVLIK, P
MONTIES, JR
机构
[1] TERACOR,TOULON,FRANCE
[2] UNIV AIX MARSEILLE 2,LRC,F-13007 MARSEILLE,FRANCE
关键词
PLATELET RETENTION TEST; TITANIUM NITRIDE; BIOMATERIALS; LEFT VENTRICULAR ASSIST DEVICE;
D O I
10.1177/039139889301600708
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In order to evaluate under dynamic circumstances the in vitro platelet adhesion induced by rigid materials such as ceramic coatings deposited on selected substrates, a new model simulating a tube has been designed. In vitro platelet adhesion was assessed with this new mode/: the material was titanium nitride (TiN) deposited on Ti6A14V (TA6V) titanium alloy by a physical vapor deposition (PVD) process. The results were compared to those obtained with complete titanium carbide (TiC) graphite tubes coated with TiN by a chemical vapor deposition (CVD) process. The difference observed (less than 25%) in favour of the new system, could be due to the better surface state of the construction materials of this system. In fact it is a systemic error. However TiN confirms its good performance as a blood-contacting biomaterial.
引用
收藏
页码:545 / 550
页数:6
相关论文
共 6 条
[1]   BIOCOMPATIBILITY OF CARBON CARBON MATERIALS - BLOOD TOLERABILITY [J].
BAQUEY, C ;
BORDENAVE, L ;
MORE, N ;
CAIX, J ;
BASSECATHALINAT, B .
BIOMATERIALS, 1989, 10 (07) :435-440
[2]  
BORDENAVE L, 1987, RBM, V6, P283
[3]  
CAZENAVE JP, 1983, ANN BIOL CLIN-PARIS, V41, P167
[4]   HEMOCOMPATIBILITY OF TITANIUM NITRIDE [J].
DION, I ;
BAQUEY, C ;
CANDELON, B ;
MONTIES, JR .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1992, 15 (10) :617-621
[5]   NEW STANNOUS AGENT KIT FOR LABELING RED BLOOD-CELLS WITH TC-99M AND ITS CLINICAL APPLICATION [J].
DUCASSOU, D ;
ARNAUD, D ;
BARDY, A ;
BEYDON, J ;
HEGESIPPE, M ;
BAQUEY, C .
BRITISH JOURNAL OF RADIOLOGY, 1976, 49 (580) :344-347
[6]   IN-111 LABELED PLATELETS - STUDIES ON PREPARATION AND EVALUATION OF IN VITRO AND IN VIVO FUNCTIONS [J].
THAKUR, ML ;
WELCH, MJ ;
JOIST, JH ;
COLEMAN, RE .
THROMBOSIS RESEARCH, 1976, 9 (04) :345-357