SCATTER IN COMPUTED RADIOGRAPHY

被引:34
作者
TUCKER, DM [1 ]
SOUTO, M [1 ]
BARNES, GT [1 ]
机构
[1] UNIV SANTIAGO DE COMPOSTELA, DEPT RADIOL, SANTIAGO, SPAIN
关键词
RADIOGRAPHY; BEDSIDE; COMPARATIVE STUDIES; COMPUTER-ASSISTED; DIGITAL; SCREENS AND FILMS;
D O I
10.1148/radiology.188.1.8511311
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The effects of scatter radiation on images obtained with storage phosphor computed radiography (CR) were evaluated. The scatter degradation factor was extended to account separately for forward scatter and backscatter. Values for the ratios of forward scatter to primary radiation (s(f)/p) and backscatter to primary radiation (S(b)/p) were measured for various x-ray tube potentials, phantom thicknesses, and field sizes for CR imaging plates (IPs). Similar measurements were made for a common screen-film system. The S(b)/p values for the IPs were 0.05-3.80. Standard CR cassettes do not have lead foil behind the IP to control backscatter; the results of this study indicate that such control is needed. The s(f)/p values were 40%-70% higher in IPs (50%-120% higher when backscatter was included) than those in the conventional screen-film system. This difference is attributed to the differences in the K-absorption edges of barium and gadolinium and the fact that scatter has lower effective energy than the primary beam as they exit the patient.
引用
收藏
页码:271 / 274
页数:4
相关论文
共 15 条
[1]  
Ballard DH, 1982, COMPUTER VISION
[2]  
Barnes G T, 1979, Med Phys, V6, P197, DOI 10.1118/1.594562
[3]  
Barnes GT, 1979, REDUCED DOSE MAMMOGR, P223
[4]  
BARNES GT, 1987, RADIOLOGY, P1
[5]   QUANTITATIVE SCATTER MEASUREMENT IN DIGITAL RADIOGRAPHY USING A PHOTOSTIMULABLE PHOSPHOR IMAGING-SYSTEM [J].
FLOYD, CE ;
LO, JY ;
CHOTAS, HG ;
RAVIN, CE .
MEDICAL PHYSICS, 1991, 18 (03) :408-413
[6]   MEASUREMENT OF SCATTER FRACTIONS IN CLINICAL BEDSIDE RADIOGRAPHY [J].
FLOYD, CE ;
BAKER, JA ;
LO, JY ;
RAVIN, CE .
RADIOLOGY, 1992, 183 (03) :857-861
[7]   QUANTITATIVE RADIOGRAPHIC IMAGING USING A PHOTOSTIMULABLE PHOSPHOR SYSTEM [J].
FLOYD, CE ;
CHOTAS, HG ;
DOBBINS, JT ;
RAVIN, CE .
MEDICAL PHYSICS, 1990, 17 (03) :454-459
[8]   LOW-DOSE DIGITAL COMPUTED RADIOGRAPHY IN PEDIATRIC CHEST IMAGING [J].
KOGUTT, MS ;
JONES, JP ;
PERKINS, DD .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (04) :775-779
[9]  
MACMAHON H, 1990, RADIOLOGY, V177, P249
[10]   SCATTERED RADIATION IN CHEST RADIOGRAPHY [J].
NIKLASON, LT ;
SORENSON, JA ;
NELSON, JA .
MEDICAL PHYSICS, 1981, 8 (05) :677-681