A NEW TECHNIQUE FOR QUANTITATION OF METAL PARTICULATES AND METAL REACTION-PRODUCTS IN TISSUES NEAR IMPLANTS

被引:16
作者
BLUMENTHAL, NC [1 ]
COSMA, V [1 ]
JAFFE, W [1 ]
STUCHIN, S [1 ]
机构
[1] HOSP JOINT DIS & MED CTR,DEPT ORTHOPAED SURG,NEW YORK,NY 10003
关键词
D O I
10.1002/jab.770050303
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Tissue specimens retrieved from four regions adjacent to hip implants during revision surgery were subjected to a novel treatment to make possible the quantitative separation of residual metal particulates and metal reaction products (metal ions and metal-protein complexes). The tissues were exposed to sodium hypochlorite solution that degraded and solubilized them, liberating metal reaction products and leaving behind metal wear particles, which were separated by centrifugation. Atomic absorption spectrophotometry was used to analyze the concentrations of the separated metal ions and wear particles. Co ion concentrations were 0.05 to 0.9 mM, Cr ion concentrations were 0.04 to 2.1 mM, and Ti ion concentrations were 0.30 to 0.60 mM. The weight of Co metal particles was 0.1 to 4.9 mg/100 mg tissue, of Cr metal particles 0.07 to 2.2 mg/100 mg tissue, and of Ti particles 0.09 to 5.2 mg/100 mg tissue; one black tissue sample contained 3333 mg Ti/100 mg tissue. No correlation was found between the concentrations of these two entities in the samples examined, probably due to the complex and varied processes creating them. The procedures discussed here will result in data that can help elucidate the separate contributions of metal reaction products and metal particulates to implant loosening. (C) 1994 John Wiley & Sons,Inc.
引用
收藏
页码:191 / 193
页数:3
相关论文
共 8 条
[1]   METALLIC WEAR IN FAILED TITANIUM-ALLOY TOTAL HIP REPLACEMENTS - A HISTOLOGICAL AND QUANTITATIVE-ANALYSIS [J].
AGINS, HJ ;
ALCOCK, NW ;
BANSAL, M ;
SALVATI, EA ;
WILSON, PD ;
PELLICCI, PM ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (03) :347-356
[2]  
[Anonymous], 1982, ANAL METHODS ATOMIC
[3]   INVITRO MODEL OF ALUMINUM-INDUCED OSTEOMALACIA - INHIBITION OF HYDROXYAPATITE FORMATION AND GROWTH [J].
BLUMENTHAL, NC ;
POSNER, AS .
CALCIFIED TISSUE INTERNATIONAL, 1984, 36 (04) :439-441
[4]   INHIBITION OF APATITE FORMATION BY TITANIUM AND VANADIUM IONS [J].
BLUMENTHAL, NC ;
COSMA, V .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH-APPLIED BIOMATERIALS, 1989, 23 (A1) :13-22
[5]  
BRIEN WW, 1992, CLIN ORTHOP RELAT R, V276, P66
[6]   SIZE OF METALLIC AND POLYETHYLENE DEBRIS PARTICLES IN FAILED CEMENTED TOTAL HIP REPLACEMENTS [J].
LEE, JM ;
SALVATI, EA ;
BETTS, F ;
DICARLO, EF ;
DOTY, SB ;
BULLOUGH, PG .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :380-384
[7]  
NAKAZONO T, 1921, J CHEM SOC JAPAN, V42, P526
[8]  
SALVATI EA, 1993, CLIN ORTHOP RELAT R, V293, P160