INVITRO TESTING OF A SIMPLY CONSTRUCTED, HIGHLY STABLE GLUCOSE SENSOR SUITABLE FOR IMPLANTATION IN DIABETIC-PATIENTS

被引:44
作者
SHAW, GW [1 ]
CLAREMONT, DJ [1 ]
PICKUP, JC [1 ]
机构
[1] UNITED MED & DENT SCH, GUYS & ST THOMAS HOSP, DIV CHEM PATHOL, LONDON SE1 9RT, ENGLAND
关键词
DIABETES-MELLITUS; GLUCOSE SENSOR; ENZYME ELECTRODE; GLUCOSE OXIDASE; POLYHYDROXYETHYL METHACRYLATE; POLYURETHANE;
D O I
10.1016/0956-5663(91)87004-U
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
We have constructed and tested in vitro a potentially implantable, needle-type amperometric enzyme electrode which is suitable for continuous monitoring of glucose concentrations in diabetic patients. The major requirements of stability during operation and ease of manufacture have been met with a sensor design which involves a simple dip-coating procedure for applying to a platinum base electrode an inner membrane of glucose oxidase immobilised in polyhydroxyethyl methacrylate (pHEMA), and an outer membrane composed of a pHEMA/polyurethane mixture. Sensors were operated at 700 mV for detection of hydrogen peroxide. Calibration curves for the sensor were linear to at least 20 mM glucose and were unaffected by a reduction in P(O2.) from 20 to 5 kPa. During continuous operation in 5 mM buffered glucose solutions in vitro, sensors suffered no significant loss of response over periods of up to 60 h. Such electrodes are, therefore, useful for development as in vivo glucose sensors.
引用
收藏
页码:401 / 406
页数:6
相关论文
共 21 条
[1]   IMMOBILIZATION OF GLUCOSE-OXIDASE IN POLY(2-HYDROXYETHYL METHACRYLATE) MEMBRANES [J].
ARICA, Y ;
HASIRCI, VN .
BIOMATERIALS, 1987, 8 (06) :489-495
[2]   APPLICATION OF CHRONIC INTRAVASCULAR BLOOD-GLUCOSE SENSOR IN DOGS [J].
ARMOUR, JC ;
LUCISANO, JY ;
MCKEAN, BD ;
GOUGH, DA .
DIABETES, 1990, 39 (12) :1519-1526
[3]  
BORETOS JW, 1981, BIOCOMPATIBILITY CLI, V2, P127
[4]   NEEDLE ENZYME ELECTRODES FOR BIOLOGICAL STUDIES [J].
CHURCHOUSE, SJ ;
BATTERSBY, CM ;
MULLEN, WH ;
VADGAMA, PM .
BIOSENSORS, 1986, 2 (06) :325-342
[5]  
CONWAY PJ, 1986, 2ND P INT M CHEM SEN, P547
[6]   TWO-DIMENSIONAL ENZYME ELECTRODE SENSOR FOR GLUCOSE [J].
GOUGH, DA ;
LUCISANO, JY ;
TSE, PHS .
ANALYTICAL CHEMISTRY, 1985, 57 (12) :2351-2357
[7]  
GOUGH DA, 1986, T AM SOC ART INT ORG, V32, P148
[8]  
Graham N., 1986, POLYM BIOMATERIALS, P170
[9]   EFFECT OF HYDROGEN PEROXIDE ON GLUCOSE OXIDASE FROM ASPERGILLUS NIGER [J].
KLEPPE, K .
BIOCHEMISTRY, 1966, 5 (01) :139-&
[10]  
KLIBANOV AM, 1979, ANAL BIOCHEM, V93, P1