RADIOSURGERY WITH UNFLATTENED 6-MV PHOTON BEAMS

被引:72
作者
OBRIEN, PF [1 ]
GILLIES, BA [1 ]
SCHWARTZ, M [1 ]
YOUNG, C [1 ]
DAVEY, P [1 ]
机构
[1] SUNNYBROOK MED CTR, DEPT NEUROSURG, TORONTO M4N 3M5, ONTARIO, CANADA
关键词
D O I
10.1118/1.596656
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
One of the major drawbacks to doing stereotactic radiosurgery with a linear accelerator is the long time required to deliver the target dose. Single fractions of 25 Gy delivered at the isocenter and at depth in the skull may require beam times in excess of 15 min for a typical linear accelerator with a maximum dose rate of 250 cGy/min in tissue. In an effort to decrease the treatment time for this technique, the flattening filter has been removed from an AECL Therac-6 linear accelerator and the characteristics of the resulting beam have been measured. Flatness is acceptable for the field sizes used with this technique and the dose rate is increased by a factor of 2.75.
引用
收藏
页码:519 / 521
页数:3
相关论文
共 11 条
[1]  
Betti OO, 1984, ACTA NEUROCHIRURGI S, V33, P385
[2]   DOSIMETRIC CONSIDERATIONS ON MULTIPLE ARC STEREOTAXIC RADIOTHERAPY [J].
CHIEREGO, G ;
MARCHETTI, C ;
AVANZO, RC ;
POZZA, F ;
COLOMBO, F .
RADIOTHERAPY AND ONCOLOGY, 1988, 12 (02) :141-152
[3]   EXTERNAL STEREOTACTIC IRRADIATION BY LINEAR-ACCELERATOR [J].
COLOMBO, F ;
BENEDETTI, A ;
POZZA, F ;
AVANZO, RC ;
MARCHETTI, C ;
CHIEREGO, G ;
ZANARDO, A .
NEUROSURGERY, 1985, 16 (02) :154-160
[4]  
GALBRAITH DM, 1985, MED PHYS, V12, P538
[5]   CEREBRAL RADIATION SURGERY USING MOVING FIELD IRRADIATION AT A LINEAR-ACCELERATOR FACILITY [J].
HARTMANN, GH ;
SCHLEGEL, W ;
STURM, V ;
KOBER, B ;
PASTYR, O ;
LORENZ, WJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (06) :1185-1192
[6]  
LARSON DA, 1990, CANCER-AM CANCER SOC, V65, P792, DOI 10.1002/1097-0142(19900201)65:3+<792::AID-CNCR2820651327>3.0.CO
[7]  
2-P
[8]   A FLATNESS AND CALIBRATION MONITOR FOR ACCELERATOR PHOTON AND ELECTRON-BEAMS [J].
MARTELL, E ;
GALBRAITH, D ;
MUNRO, P ;
RAWLINSON, JA ;
TAYLOR, WB .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1986, 12 (02) :271-275
[9]   DOSE DISTRIBUTIONS IN DYNAMIC STEREOTAXIC RADIOSURGERY [J].
PIKE, B ;
PODGORSAK, EB ;
PETERS, TM ;
PLA, C .
MEDICAL PHYSICS, 1987, 14 (05) :780-789
[10]   DYNAMIC STEREOTACTIC RADIOSURGERY [J].
PODGORSAK, EB ;
OLIVIER, A ;
PLA, M ;
LEFEBVRE, PY ;
HAZEL, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (01) :115-126