COMPARISON OF THEOPHYLLINE BLOOD-LEVELS MEASURED BY THE STANDARD TDX ASSAY AND A NEW PATIENT-SIDE IMMUNOASSAY CARTRIDGE SYSTEM

被引:3
作者
KLOTZ, U
机构
[1] Dr. Margarete Fischer-Bosch-Institute of Clinical Pharmacology, Stuttgart
关键词
THEOPHYLLINE; IMMUNOASSAYS; TDX; CARTRIDGE SYSTEM;
D O I
10.1097/00007691-199310000-00017
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Plasma steady-state concentrations of theophylline were measured by the standard TDx assay and a novel patient-side immunoassay system with a single-use cartridge (Biotrack 516) in 100 samples from 93 patients. Intra- and interassay coefficients of variation for the new cartridge technique averaged 4.3 and 9.8%, respectively. In the concentration range of 2 to 40 mug/ml, there was a linear correlation (r = 0.96, p < 0.0001) between both assays [slope of 0.86; y-intercept (TDx values) = -0.14 mug/ml], indicating that the two methods can be used interchangeably. Since the completely automated cartridge system is very simple and extremely rapid, this technique offers some advantages for routine clinical use, especially under ambulatory conditions.
引用
收藏
页码:462 / 464
页数:3
相关论文
共 6 条
[1]   ABC OF MONITORING DRUG-THERAPY - THEOPYLLINE [J].
ARONSON, JK ;
HARDMAN, M ;
REYNOLDS, DJM .
BRITISH MEDICAL JOURNAL, 1992, 305 (6865) :1355-1358
[2]   THERAPEUTIC MONITORING OF THEOPHYLLINE - RATIONALE AND CURRENT STATUS [J].
BIERMAN, CW ;
WILLIAMS, PV .
CLINICAL PHARMACOKINETICS, 1989, 17 (06) :377-384
[3]   AN UPDATED COMPARISON OF DRUG DOSING METHODS .2. THEOPHYLLINE [J].
ERDMAN, SM ;
RODVOLD, KA ;
PRYKA, RD .
CLINICAL PHARMACOKINETICS, 1991, 20 (04) :280-292
[4]  
GIBBONS I, 1989, CLIN CHEM, V35, P1869
[5]  
UEMATSU T, 1985, PRACT CARDIOL, V11, P95
[6]   EVALUATION OF CHROMATOGRAPHIC AND KIT IMMUNOASSAY TECHNIQUES FOR THE MEASUREMENT OF THEOPHYLLINE IN SERUM - A STUDY BASED ON EXTERNAL QUALITY ASSURANCE MEASUREMENTS [J].
WILSON, JF ;
WILLIAMS, J ;
TSANACLIS, LM ;
TEDSTONE, JE ;
RICHENS, A .
THERAPEUTIC DRUG MONITORING, 1988, 10 (04) :438-445