DETECTION OF COCAINE USING THE FLOW IMMUNOSENSOR

被引:30
作者
OGERT, RA [1 ]
KUSTERBECK, AW [1 ]
WEMHOFF, GA [1 ]
BURKE, R [1 ]
LIGLER, FS [1 ]
机构
[1] USN,RES LAB,CTR BIOMOLEC SCI & ENGN,CODE 6090,WASHINGTON,DC 20375
关键词
COCAINE DETECTION; IMMUNOSENSOR;
D O I
10.1080/00032719208020071
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A continuous flow immunosensor has been designed for the detection of cocaine in aqueous samples. The continuous flow immunosensor relies on the displacement of fluorophore-labeled antigen from immobilized monoclonal antibody. The sensitivity and accuracy of the flow immunosensor were investigated while varying the parameters of immobilized antibody density, flow rate, amount of antibody-coated Sepharose used in each column, and the saturation of antibody binding sites with fluorophore-labeled antigen. Using a low density of immobilized anti-benzoylecgonine antibody, as little as 5 ng/ml cocaine could be detected. Small amounts of antibody-coated Sepharose could be used repeatedly and the lifetime of the column was proportional to the amount of Sepharose used. Results were obtained in less than a minute and cross-reactivity against various other drugs was negligible.