DISASSEMBLY OF A ONE-PIECE METAL-BACKED ACETABULAR COMPONENT - A CASE-REPORT

被引:12
作者
BEAVER, RJ [1 ]
SCHEMITSCH, EH [1 ]
GROSS, AE [1 ]
机构
[1] UNIV TORONTO,MT SINAI HOSP,DIV ORTHOPAED SURG,600 UNIV AVE,SUITE 476A,TORONTO M5G 1X5,ONTARIO,CANADA
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1991年 / 73卷 / 06期
关键词
D O I
10.1302/0301-620X.73B6.1955434
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a case of late dissociation of a one-piece metal-backed acetabular component, associated with a prolific soft-tissue reaction to the polyethylene debris. The polyethylene liner was not visible on the radiographs. The metal-backed shell could not be removed because of bone ingrowth. We recommend that modular systems be utilised in cementless arthroplasty of the hip and that radiographic markers be incorporated in the polyethylene of the acetabular cup.
引用
收藏
页码:908 / 910
页数:3
相关论文
共 6 条
[1]
Bueche M J, 1989, J Arthroplasty, V4, P39, DOI 10.1016/S0883-5403(89)80052-X
[2]
Ferenz C C, 1988, J Arthroplasty, V3, P201
[3]
DISASSEMBLY OF A MODULAR ACETABULAR COMPONENT OF A TOTAL HIP-REPLACEMENT ARTHROPLASTY - A CASE-REPORT [J].
KITZIGER, KJ ;
DELEE, JC ;
EVANS, JA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :621-623
[4]
DISASSEMBLY OF A MODULAR FEMORAL COMPONENT DURING CLOSED REDUCTION OF THE DISLOCATED FEMORAL COMPONENT - A CASE-REPORT [J].
PELLICCI, PM ;
HAAS, SB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :619-620
[5]
ACETABULAR CUP DISLOCATION - A NEW COMPLICATION OF TOTAL JOINT ARTHROPLASTY [J].
WILSON, AJ ;
MONSEES, B ;
BLAIR, VP .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (01) :133-134
[6]
DISASSEMBLY OF A MODULAR FEMORAL PROSTHESIS AFTER DISLOCATION OF THE FEMORAL COMPONENT - A CASE-REPORT [J].
WOOLSON, ST ;
POTTORFF, GT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :624-625