CABLE COUPLING OF DETECTOR AND PREAMPLIFIER AS AN ALTERNATIVE DESIGN OF RADIATION-HARD ELECTRONICS

被引:6
作者
AKOPYAN, M
CHEN, M
FEDYAKIN, N
KRASNOKUTSKY, R
SHUVALOV, R
SUSHKOV, V
ZHANG, X
机构
[1] RUSSIAN ACAD SCI,INST NUCL RES,MOSCOW,RUSSIA
[2] MIT,CAMBRIDGE,MA 02139
关键词
D O I
10.1016/0168-9002(93)90577-5
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
It is shown that cable coupling between a detector of large capacitance and a fast amplifier does not degrade noise performance. The dependences of equivalent noise charge (ENC) on cable length, detector capacitance and output pulse duration are presented. There is a fairly good agreement between experiment and calculation.