PREDICTABILITY OF SNR AND READER PREFERENCE IN CLINICAL MR IMAGING

被引:16
作者
OWEN, RS
WEHRLI, FW
机构
[1] The Hospital of the University of Pennsylvania, Philadelphia
关键词
Image quality; Magnetic resonance; Magnetic resonance imaging; Signal-to-noise ratio (SNR);
D O I
10.1016/0730-725X(90)90009-Q
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Fifty-four independent scans were performed in two volunteers covering one anatomic region in each (the brain and knee) with the purpose of ascertaining the agreement between predicted and measured signal-to-noise ratios (SNR). Systematically varied parameters were number of excitations (NEX), field of view (FOV), section thickness (dz), and the number of phase-encoding steps (Ny). Correlation coefficients of measured versus predicted SNR were 0.82 and 0.86, respectively, in the anatomies studied. Significantly improved correlations were found for data subpopulations in which NEX was held constant. To assess the criteria guiding reader preference, a blinded study was performed in which radiologists were asked to rate images from least to most desirable. In order to quantitatively determine the criteria for reader preference, plots of mean rating versus SNR, voxel volume, and an image quality index [IQI = SNR/(voxel volume)] were performed. The latter was found to be a better predictor of reader preference than either SNR or spatial resolution alone. The data suggests T1-weighted scan protocols yielding SNR of approximately 20 are preferable with any excess SNR being traded for smaller voxel size or shorter scan times. © 1990.
引用
收藏
页码:737 / 745
页数:9
相关论文
共 6 条
  • [1] Bradley, Kortman, Crues, Central nervous system high-resolution magnetic resonance imaging: effects of increasing spatial resolution on resolving power, Radiology, 156, pp. 93-98, (1985)
  • [2] Bradley, Tsuruda, MR sequence parameter optimization: An algorithmic approach, AJR, 149, pp. 815-823, (1987)
  • [3] Edelstein, Glover, Hardy, Redington, The intrinsic signal-to-noise ratio in NMR imaging, Magn. Res. Med., 3, pp. 604-618, (1986)
  • [4] Prost, Wehrli, Drayer, Et al., SAR reduced pulse sequences, Magn. Reson. Imaging, 6, pp. 125-130, (1988)
  • [5] Wehrli, Signal-to-noise and contrast in MR imaging, NMR in Medicine. The Instrumentation and Clinical Application, (1986)
  • [6] Kneeland, Shimakawa, Wehrli, Effects of intersection spacing on MR image contrast and study time, Radiology, 158, pp. 819-822, (1986)