IMPORTANT CONSIDERATIONS IN MEASUREMENT OF ATTRACTIVE FORCE ON METALLIC IMPLANTS IN MR IMAGERS

被引:47
作者
KAGETSU, NJ
LITT, AW
机构
[1] Department of Radiology, New York University, Medical Center, New York, NY 10016
关键词
FOREIGN BODIES; MAGNETIC RESONANCE (MR); PHYSICS; SAFETY; RADIOLOGY AND RADIOLOGISTS; IATROGENIC INJURY;
D O I
10.1148/radiology.179.2.2014301
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Almost all reported measurements of the attractive force exerted on metallic implants in magnetic fields of magnetic resonance imagers have been obtained at the portal. This study was done to determine whether the maximum force on an implant is at the portal. Horizontal attractive forces on needles were measured along the axis of the bore at various distances from the portal of a 0.3-T vertical-field magnet and 0.5-and 1.5-T solenoid magnets. Upward vertical forces along the top of the bore were also measured. The horizontal forces 40 cm in from the portals of the 0.5- and 1.5-T magnets were 70%-80% greater than the horizontal forces at the portal. The upward vertical force at the top of the 0.3-T magnet bore, 63 cm in from the portal, was 20 times the horizontal force at the portal and was twice the maximum force measured in the 0.5-T magnet. The maximum force was not at the portal.
引用
收藏
页码:505 / 508
页数:4
相关论文
共 16 条
  • [1] APPLEBAUM EL, 1985, ARCH OTOLARYNGOL, V111, P820
  • [2] MAGNETIC-RESONANCE IMAGER SURVEY, JUNE 1985
    BRADLEY, WG
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1985, 9 (06) : 1155 - 1165
  • [3] MAGNETIC-FIELD EFFECTS ON SURGICAL LIGATION CLIPS
    BROWN, MA
    CARDEN, JA
    COLEMAN, RE
    MCKINNEY, R
    SPICER, LD
    [J]. MAGNETIC RESONANCE IMAGING, 1987, 5 (06) : 443 - 453
  • [4] ANEURYSM CLIP MOTION DURING MAGNETIC-RESONANCE IMAGING - INVIVO EXPERIMENTAL-STUDY WITH METALLURGICAL FACTOR-ANALYSIS
    DUJOVNY, M
    KOSSOVSKY, N
    KOSSOWSKY, R
    VALDIVIA, R
    SUK, JS
    DIAZ, FG
    BERMAN, SK
    CLEARY, W
    [J]. NEUROSURGERY, 1985, 17 (04) : 543 - 548
  • [5] THE EFFECTS OF MAGNETIC-RESONANCE IMAGING ON DIFFERENT TYPES OF MICROSURGICAL CLIPS
    KEAN, DM
    WORTHINGTON, BS
    FIRTH, JL
    HAWKES, RC
    [J]. JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1985, 48 (03) : 286 - 287
  • [6] KELLY WM, 1986, AM J NEURORADIOL, V7, P243
  • [7] MESGARZADEH M, 1985, SKELETAL RADIOL, V14, P205, DOI 10.1007/BF00355565
  • [8] POTENTIAL HAZARDS AND ARTIFACTS OF FERROMAGNETIC AND NONFERROMAGNETIC SURGICAL AND DENTAL MATERIALS AND DEVICES IN NUCLEAR MAGNETIC-RESONANCE IMAGING
    NEW, PFJ
    ROSEN, BR
    BRADY, TJ
    BUONANNO, FS
    KISTLER, JP
    BURT, CT
    HINSHAW, WS
    NEWHOUSE, JH
    POHOST, GM
    TAVERAS, JM
    [J]. RADIOLOGY, 1983, 147 (01) : 139 - 148
  • [9] Purcell E.M., 1965, ELECTRICITY MAGNETIS, V2
  • [10] MR IMAGING OF METALLIC IMPLANTS AND MATERIALS - A COMPILATION OF THE LITERATURE
    SHELLOCK, FG
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 1988, 151 (04) : 811 - 814