HIGH-RESOLUTION MICROCOIL H-1-NMR FOR MASS-LIMITED, NANOLITER-VOLUME SAMPLES

被引:432
作者
OLSON, DL
PECK, TL
WEBB, AG
MAGIN, RL
SWEEDLER, JV
机构
[1] UNIV ILLINOIS, BECKMAN INST, MAGNET RESONANCE ENGN LAB, URBANA, IL 61801 USA
[2] UNIV ILLINOIS, DEPT CHEM, URBANA, IL 61801 USA
[3] UNIV ILLINOIS, DEPT ELECT & COMP ENGN, URBANA, IL 61801 USA
关键词
D O I
10.1126/science.270.5244.1967
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
High-resolution, proton nuclear magnetic resonance (NMR) spectra of 5-nanoliter samples have been obtained with much higher mass sensitivity [signal-to-noise ratio (S/N) per micromole] than with traditional methods. Arginine and sucrose show a mean sensitivity enhancement of 130 compared to 278-microliter samples run in a 5-millimeter tube in a conventional, commercial probe. This can reduce data acquisition time by a factor of >16,000 or reduce the needed sample mass by a factor of about 130. A linewidth of 0.6 hertz was achieved on a 360-megahertz spectrometer by matching the magnetic susceptibility of the medium that surrounds the detection cell to of the copper coil. For sucrose, the limit of detection (defined at S/N = 3) was 19 nanograms (56 picomoles) for a 1-minute data acquisition. This technique should prove useful with mass-limited samples and for use as a detector in capillary separations.