Pulse compression techniques that are capable of producing a large signal-to-noise (SNR) enhancement, have been used successfully in many different fields. For medical applications, frequency-dependent attenuation in soft tissue can limit the usefulness of this method. In the paper, this issue is examined through model-simulation studies. Frequency-modulation (FM) chirp, considered in the study, is just one form of pulse coding technique. Pulse propagation effects in soft tissue are modelled as a linear zero phase filter. A method to perform simulations and estimate the effective time-bandwidth product K is outlined. K describes the SNR enhancement attainable under limitations imposed by the soft-tissue medium. An effective time-bandwidth product is evaluated as a function of soft-tissue linear attenuation coefficient alpha(o), scatterer depth z and the bandwidth of the interrogating FM pulse, under realistic conditions. Results indicate that, under certain conditions, K can be significantly lower than its expected value in a non-attenuating medium. It is argued that although limitations exist, pulse compression techniques can still be used to improve resolution or increase penetrational depth. The real advantage over conventional short-pulse imaging comes from the possibility that these improvements can be accomplished without increasing the peak intensity of the interrogating pulse above any threshold levels set by possible bio-effect considerations.
引用
收藏
页码:181 / 188
页数:8
相关论文
共 19 条
[1]   GAUGING THE LIKELIHOOD OF CAVITATION FROM SHORT-PULSE, LOW-DUTY CYCLE DIAGNOSTIC ULTRASOUND [J].
APFEL, RE ;
HOLLAND, CK .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (02) :179-185
[2]  
APFEL RE, 1982, BRIT J CANCER, V45, P140
[3]  
CHAPELON JY, 1988, PROGR MED IMAGING, pCH6
[4]   THE OUTPUT OF PULSE-ECHO ULTRASOUND EQUIPMENT - A SURVEY OF POWERS, PRESSURES AND INTENSITIES [J].
DUCK, FA ;
STARRITT, HC ;
AINDOW, JD ;
PERKINS, MA ;
HAWKINS, AJ .
BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (694) :989-1001
[5]  
GROSS SA, 1980, J ACOUST SOC AM, V68, P93
[6]  
GURUMURTHY KV, 1981, ULTRSON IMAG, V4, P355
[7]   ULTIMATE LIMITS IN ULTRASONIC-IMAGING RESOLUTION [J].
HARRIS, RA ;
FOLLETT, DH ;
HALLIWELL, M ;
WELLS, PNT .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (06) :547-558
[8]   WHAT MIGHT ECHOGRAPHY LEARN FROM IMAGE SCIENCE [J].
HILL, CR ;
BAMBER, JC ;
CRAWFORD, DC ;
LOWE, HJ ;
WEBB, S .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (06) :559-575
[9]   SIGNAL-PROCESSING OF BROAD-BAND PULSED ULTRASOUND - MEASUREMENT OF ATTENUATION OF SOFT BIOLOGICAL TISSUES [J].
KAK, AC ;
DINES, KA .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1978, 25 (04) :321-344
[10]   THE THEORY AND DESIGN OF CHIRP RADARS [J].
KLAUDER, JR ;
PRICE, AC ;
DARLINGTON, S ;
ALBERSHEIM, WJ .
BELL SYSTEM TECHNICAL JOURNAL, 1960, 39 (04) :745-808