DOPPLER HEMODYNAMIC ASSESSMENT OF CLINICALLY AND ECHOCARDIOGRAPHICALLY NORMAL MITRAL AND AORTIC ALLCARBON VALVE PROSTHESES

被引:10
作者
BADANO, L
BERTOLI, D
ASTENGO, D
CARRATINO, L
DEGAETANO, G
PASSERONE, GC
CAMERIERI, A
GRAMENZI, S
MAGAJA, O
FAZZINI, L
PAPAGNA, D
BIANCHI, F
BARBERIS, L
LUCATTI, A
机构
[1] OSPED SAN BARTOLOMEO,SERV CARDIOL,SARZANA,ITALY
[2] OSPED CELESIA,SERV CARDIOL,GENOA,ITALY
[3] UNIV GENOA,CATTEDRA CARDIOCHIRURG,GENOA,ITALY
[4] OSPED SAN GIOVANNI BATTISTA TORINO,SERV CARDIOL,GENOA,ITALY
[5] OSPED CIVILE,SERV CARDIOL,LA SPEZIA,ITALY
[6] OSPED SP ARENA,SERV CARDIOL,GENOA,ITALY
[7] OSPED SAN MARTINO GENOVA,SERV CARDIOL,I-16132 GENOA,ITALY
[8] OSPED RIUNITI,SERV CARDIOL,LAVAGNA,ITALY
关键词
ALLCARBON; PROSTHETIC VALVES; DOPPLER ECHOCARDIOGRAPHY;
D O I
10.1093/eurheartj/14.12.1602
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Doppler echocardiographic characteristics of normally functioning Allcarbon prostheses were studied in 149 consecutive patients with 157 valves in the mitral (n=73) and aortic (n=84) positions whose function was considered normal by clinical and echocardiographic evaluation. In the mitral position, the mean gradient and the effective mitral orifice area were not significantly different in either the 25-mm or the 31-mm size valves (from 5±1 to 4±1 mmHg and from 2.2±0.6 to 2.8±0.9 cm2, respectively; P=ns for both). Conversely, peak gradient was significantly and inversely correlated to actual orifice area (r=-0.70; P<0.0006), decreasing from 15±3 mmHg in the 25-mm size valve to 9±1 mmHg in the 31-mm size.In the aortic position, the mean gradient was 29±8 mmHg in the 19-mm size valve; it decreased to 8±2 mmHg in the 29-mm size. Effective prosthetic aortic valve area, calculated using the continuity equation, ranged between 0.9±0.1 cm2 for the 19-mm size valve to 4.1±0.7 cm2 for the 29-mm size. By analysis of variance, effective prosthetic aortic valve area differentiated various valve sizes (F=25.3; P<0.0001) better than peak (F=5.34; P=0.012) or mean (F=4.34; P=0.0052) gradients alone, and it correlated better with actual orifice area (r=0.89, r=-0.70 and r=-0.65, respectively). This study provides the normal range for Doppler haemodynamic characteristics of the various sizes of the Allcarbon valve in the mitral and aortic positions so that prosthetic malfunction can be identified. © 1993 The Europen Society of Cardiology.
引用
收藏
页码:1602 / 1609
页数:8
相关论文
共 31 条
[1]   CONTINUOUS WAVE DOPPLER ECHOCARDIOGRAPHIC MEASUREMENT OF PROSTHETIC VALVE GRADIENTS - A SIMULTANEOUS DOPPLER-CATHETER CORRELATIVE STUDY [J].
BURSTOW, DJ ;
NISHIMURA, RA ;
BAILEY, KR ;
REEDER, GS ;
HOLMES, DR ;
SEWARD, JB ;
TAJIK, AJ .
CIRCULATION, 1989, 80 (03) :504-514
[2]   DOPPLER ECHOCARDIOGRAPHIC ASSESSMENT OF THE ST-JUDE MEDICAL PROSTHETIC VALVE IN THE AORTIC POSITION USING THE CONTINUITY EQUATION [J].
CHAFIZADEH, ER ;
ZOGHBI, WA .
CIRCULATION, 1991, 83 (01) :213-223
[3]   ECHOCARDIOGRAPHIC DESCRIPTION OF THE CARBOMEDICS BILEAFLET PROSTHETIC HEART-VALVE [J].
CHAMBERS, J ;
CROSS, J ;
DEVERALL, P ;
SOWTON, E .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1993, 21 (02) :398-405
[4]   VALIDATION AND APPLICATIONS OF MITRAL PROSTHETIC VALVULAR AREAS CALCULATED BY DOPPLER ECHOCARDIOGRAPHY [J].
DUMESNIL, JG ;
HONOS, GN ;
LEMIEUX, M ;
BEAUCHEMIN, J .
AMERICAN JOURNAL OF CARDIOLOGY, 1990, 65 (22) :1443-1448
[5]   HEMODYNAMIC EVALUATION OF PORCINE BIOPROSTHESES IN THE MITRAL POSITION BY DOPPLER ECHOCARDIOGRAPHY [J].
FAWZY, ME ;
HALIM, M ;
ZIADY, G ;
MERCER, E ;
PHILLIPS, R ;
ANDAYA, W .
AMERICAN JOURNAL OF CARDIOLOGY, 1987, 59 (06) :643-646
[6]  
FOX RW, 1978, INTRO FLUID MECHANIC, P449
[7]  
Gabbay S, 1979, Circulation, V60, P62
[8]  
GABBAY S, 1978, J THORAC CARDIOV SUR, V76, P771
[9]  
GIOIA G, 1988, J ITAL CARDIOL, V18, P213
[10]   BILEAFLET, TILTING DISK AND PORCINE AORTIC-VALVE SUBSTITUTES - INVIVO HYDRODYNAMIC CHARACTERISTICS [J].
GRAY, RJ ;
CHAUX, A ;
MATLOFF, JM ;
DEROBERTIS, M ;
RAYMOND, M ;
STEWART, M ;
YOGANATHAN, A .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1984, 3 (02) :321-327