REGIONAL MYOCARDIAL BLOOD-FLOW WITH A REPERFUSION CATHETER AND AN AUTOPERFUSION BALLOON CATHETER DURING TOTAL CORONARY-OCCLUSION

被引:16
作者
CHRISTENSEN, CW
LASSAR, TA
DALEY, LC
RIEDER, MA
SCHMIDT, DH
机构
关键词
D O I
10.1016/S0002-8703(05)80011-0
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We investigated the ability of two new coronary perfusion catheters to maintain regional myocardial blood flow throughout a 90-minute period of occlusion. In 21 dogs (group I=total occlusion control; group II=reperfusion catheter; group III=autoperfusion balloon catheter) we studied regional blood flow, distal coronary perfusion pressure, infarct size, and red blood cell hemolysis after placement of either catheter into the left anterior descending coronary artery. Regional (microsphere) blood flow showed a reduction in transmural blood flow during occlusion in comparison to baseline values (1.07±0.12 to 0.81±0.11 and 1.01±0.16 to 0.73±0.08 ml/min subendocardial perfusion for groups II and III, respectively). Comparable changes in blood flow were observed in the subepicardial and midmyocardial regions. Distal, coronary perfusion pressures were reduced by 26% and 28% for groups II and III, respectively. Both catheters prevented significant infarction and maintained adequate regional myocardial blood flow throughout the 90-minute period of occlusion without significant complications of clotting or destruction of erythrocytes. © 1990.
引用
收藏
页码:242 / 248
页数:7
相关论文
共 10 条
[1]   DISTAL CORONARY-ARTERY PERFUSION DURING PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
ANDERSON, HV ;
LEIMGRUBER, PP ;
ROUBIN, GS ;
NELSON, DL ;
GRUENTZIG, AR .
AMERICAN HEART JOURNAL, 1985, 110 (04) :720-726
[2]   INADEQUATE SUBENDOCARDIAL OXYGEN DELIVERY DURING PERFLUOROCARBON PERFUSION IN A CANINE MODEL OF ISCHEMIA [J].
CHRISTENSEN, CW ;
REEVES, WC ;
LASSAR, TA ;
SCHMIDT, DH .
AMERICAN HEART JOURNAL, 1988, 115 (01) :30-37
[3]   NEW BALLOON CATHETER FOR PROLONGED PERCUTANEOUS TRANS-LUMINAL CORONARY ANGIOPLASTY AND BYPASS FLOW IN OCCLUDED VESSELS [J].
ERBEL, R ;
CLAS, W ;
BUSCH, U ;
VONSEELEN, W ;
BRENNECKE, R ;
BLOMER, H ;
MEYER, J .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1986, 12 (02) :116-123
[4]   CATHETER REPERFUSION TO ALLOW OPTIMAL CORONARY-BYPASS GRAFTING FOLLOWING FAILED TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
FERGUSON, TB ;
HINOHARA, T ;
SIMPSON, J ;
STACK, RS ;
WECHSLER, AS .
ANNALS OF THORACIC SURGERY, 1986, 42 (04) :399-405
[5]   BLOOD-FLOW MEASUREMENTS WITH RADIONUCLIDE-LABELED PARTICLES [J].
HEYMANN, MA ;
PAYNE, BD ;
HOFFMAN, JIE ;
RUDOLPH, AM .
PROGRESS IN CARDIOVASCULAR DISEASES, 1977, 20 (01) :55-79
[6]   TRANS-LUMINAL INTRACORONARY REPERFUSION CATHETER - A DEVICE TO MAINTAIN CORONARY PERFUSION BETWEEN FAILED CORONARY ANGIOPLASTY AND EMERGENCY CORONARY-BYPASS SURGERY [J].
HINOHARA, T ;
SIMPSON, JB ;
PHILLIPS, HR ;
STACK, RS .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1988, 11 (05) :977-982
[7]   PROLONGED APPLICATION OF PRESSURE IN TRANS-LUMINAL CORONARY ANGIOPLASTY [J].
KALTENBACH, M ;
BEYER, J ;
WALTER, S ;
KLEPZIG, H ;
SCHMIDTS, L .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1984, 10 (03) :213-219
[8]   MYOCARDIAL PROTECTION DURING CORONARY ANGIOPLASTY WITH AN AUTOPERFUSION BALLOON CATHETER IN HUMANS [J].
QUIGLEY, PJ ;
HINOHARA, T ;
PHILLIPS, HR ;
PETER, RH ;
BEHAR, VS ;
KONG, YH ;
SIMONTON, CA ;
PEREZ, JA ;
STACK, RS .
CIRCULATION, 1988, 78 (05) :1128-1134
[9]   USEFULNESS OF INTRACORONARY INFUSION OF FLUOROCARBON DISTAL TO PROLONGED CORONARY-OCCLUSION BY ANGIOPLASTY BALLOON IN DOGS [J].
ROBERTS, CS ;
ANDERSON, HV ;
CARBONI, AA ;
JUSTICZ, AGS ;
LEIMGRUBER, PP ;
KLONER, RA ;
GRUENTZIG, AR .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (13) :1202-1205
[10]   PRESERVATION OF DISTAL CORONARY PERFUSION DURING PROLONGED BALLOON INFLATION WITH AN AUTOPERFUSION ANGIOPLASTY CATHETER [J].
TURI, ZG ;
CAMPBELL, CA ;
GOTTIMUKKALA, MV ;
KLONER, RA .
CIRCULATION, 1987, 75 (06) :1273-1280