FALSE-POSITIVE LUNG IMAGING - INADVERTENT INJECTION INTO A PULMONARY-ARTERY CATHETER

被引:3
作者
BRACHMAN, M
TANASESCU, D
RAMANNA, L
BERMAN, D
WAXMAN, A
机构
关键词
D O I
10.1097/00003072-197910000-00005
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A case of ventilation-perfusion mismatch seen postoperatively in a patient with chest pain is reported. There was absence of perfusion to the right lung, with relatively normal ventilation. The study was initially interpreted as indicating a high probability of pulmonary embolus. It was then discovered that the injection had been inadvertently made into a Swan-Ganz catheter, with its tip in the left main pulmonary artery. The mismatch was therefore iatrogenic and not related to pulmonary embolus. The false-positive lung imaging which resulted has not been previously reported in the literature.