DIRECT MEASUREMENT OF TEMPERATURE-CHANGES WITHIN FREEZE-FRACTURE SPECIMENS DURING RAPID QUENCHING IN LIQUID COOLANTS

被引:166
作者
COSTELLO, MJ
CORLESS, JM
机构
[1] Department of Anatomy, Duke University Medical Center, Durham, North Carolina, 27710
关键词
D O I
10.1111/j.1365-2818.1978.tb01151.x
中图分类号
TH742 [显微镜];
学科分类号
摘要
We have evaluated the cooling rates of specimens mounted in a variety of freeze‐fracture holders when plunged into a series of liquid coolants. These rates were measured using miniature thermocouples placed within the mounted specimens. The most rapid cooling rates were obtained using propane at 83 K as the coolant. When mounted on a newly devised ‘copper sandwich’ holder, specimen cooling rates in excess of 4500 K/s have been recorded. A simple guillotine‐like device for quenching freeze‐fracture specimens under reproducible conditions is presented. 1978 Blackwell Science Ltd
引用
收藏
页码:17 / 37
页数:21
相关论文
共 27 条
[1]   CORRELATED X-RAY-DIFFRACTION AND FREEZE-FRACTURE STUDIES ON MEMBRANE MODEL SYSTEMS PERTURBATIONS INDUCED BY FREEZE-FRACTURE PREPARATIVE PROCEDURES [J].
COSTELLO, MJ ;
GULIKKRZYWICKI, T .
BIOCHIMICA ET BIOPHYSICA ACTA, 1976, 455 (02) :412-432
[2]  
COWLEY C. W., 1961, BIODYNAMICA, V8, P317
[3]  
ESCAIG J, 1977, CR ACAD SCI D NAT, V284, P2289
[4]   USE OF LOW-TEMPERATURE X-RAY-DIFFRACTION TO EVALUATE FREEZING METHODS USED IN FREEZE-FRACTURE ELECTRON-MICROSCOPY [J].
GULIKKRZYWICKI, T ;
COSTELLO, MJ .
JOURNAL OF MICROSCOPY, 1978, 112 (JAN) :103-113
[5]  
HEUSER JE, 1975, COLD SPRING HARB SYM, V40, P17
[7]  
LUYET B., 1962, BIODYNAMICA, V9, P21
[8]  
LUYET B., 1961, BIODYNAMICA, V8, P331
[9]  
Luyet B., 1951, REFRIG ENG, V59, P1191
[10]  
Luyet B.J., 1960, 4 ANN M BIOPH SOC, P31