THEORETICAL RESOLUTION OF COMPUTED TOMOGRAPHY SYSTEMS

被引:39
作者
GLOVER, GH
EISNER, RL
机构
[1] Applied Science and Diagnostic Laboratory, General Electric Medical Systems, Milwaukee, WI, 53201
关键词
Computed tomography; Modulation transfer function; Point response function; Resolution;
D O I
10.1097/00004728-197902000-00014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A simple analytic model is presented which is used to calculate the impulse response of a computed tomography system for parametric variations of the geometry. The model is useful for fan beam as well as parallel beam geometries. Results show that wide-aperture detector systems have intrinsic resolution limitations that are not mitigated by oversampling. Furthermore, the optimum focal spot size of the X-ray source is found to be dependent on the detector geometry. © 1979 Raven Press, New York.