A COMBINED TDDA-PVC PH AND REFERENCE ELECTRODE FOR USE IN THE UPPER SMALL-INTESTINE

被引:26
作者
MCEWAN, GTA
LUCAS, ML
DENVIR, M
RAJ, M
MCCOLL, KEL
RUSSELL, RI
MATHAN, VI
机构
[1] UNIV GLASGOW,INST PHYSIOL,GLASGOW G12 8QQ,SCOTLAND
[2] WESTERN INFIRM & ASSOCIATED HOSP,GLASGOW G4 0SF,SCOTLAND
[3] ROYAL INFIRM,GLASGOW G31 2ER,SCOTLAND
[4] CMC HOSP,DEPT GASTROENTEROL,WELLCOME RES UNIT,VELLORE,TAMIL NADU,INDIA
基金
英国惠康基金;
关键词
D O I
10.3109/03091909009028758
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Incorporation of a ligand into polyvinylchloride allows the production of ion selective electrodes that are soft-bodied and disposable. With features make them especially suitable for clinical, particularly gastroenterological, investigations. We report here on the construction of a combined reference and pH electrode suitable for use at jejunal biopsy. With this type of pH electrode, the pH of the mucosal surface of the jejunum in UK and in Indian subjects, without evidence of upper gastrointestinal disease, was about pH 6.0. This was almost identical to previous values measured using a separate reference electrode. Both the polymer electrode and a suitable data logger can be conveniently produced in the laboratory and compare favourably with commercially available systems, provided the range of pH likely to be encountered is within the operating range of the incorporated ligand. © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
引用
收藏
页码:16 / 20
页数:5
相关论文
共 11 条
[1]   TECHNICAL ASPECTS OF INTRALUMINAL PH-METRY IN MAN - CURRENT STATUS AND RECOMMENDATIONS [J].
EMDE, C ;
GARNER, A ;
BLUM, AL .
GUT, 1987, 28 (09) :1177-1188
[2]  
HOGBEN CAM, 1959, J PHARMACOL EXP THER, V125, P275
[3]   DETERMINATION OF ACID SURFACE PH INVIVO IN RAT PROXIMAL JEJUNUM [J].
LUCAS, M .
GUT, 1983, 24 (08) :734-739
[4]   JEJUNAL SURFACE PH MEASUREMENTS IN TROPICAL-SPRUE [J].
LUCAS, ML ;
MATHAN, VI .
TRANSACTIONS OF THE ROYAL SOCIETY OF TROPICAL MEDICINE AND HYGIENE, 1989, 83 (01) :138-142
[5]   ACID MICROCLIMATE IN CELIAC AND CROHNS-DISEASE - MODEL FOR FOLATE MALABSORPTION [J].
LUCAS, ML ;
COOPER, BT ;
LEI, FH ;
JOHNSON, IT ;
HOLMES, GKT ;
BLAIR, JA ;
COOKE, WT .
GUT, 1978, 19 (08) :735-742
[6]  
McCloy R. F., 1970, Clinical Physics and Physiological Measurement, V1, P151, DOI 10.1088/0143-0815/1/2/006
[7]   THE EFFECT OF ESCHERICHIA-COLI STA ENTERO-TOXIN AND OTHER SECRETAGOGUES ON MUCOSAL SURFACE PH OF RAT SMALL-INTESTINE INVIVO [J].
MCEWAN, GTA ;
DANIEL, H ;
FETT, C ;
BURGESS, MN ;
LUCAS, ML .
PROCEEDINGS OF THE ROYAL SOCIETY SERIES B-BIOLOGICAL SCIENCES, 1988, 234 (1275) :219-237
[8]   DESIGN OF NEUTRAL HYDROGEN-ION CARRIERS FOR SOLVENT POLYMERIC MEMBRANE ELECTRODES OF SELECTED PH RANGE [J].
OESCH, U ;
BRZOZKA, Z ;
XU, AP ;
RUSTERHOLZ, B ;
SUTER, G ;
PHAM, HV ;
WELTI, DH ;
AMMANN, D ;
PRETSCH, E ;
SIMON, W .
ANALYTICAL CHEMISTRY, 1986, 58 (11) :2285-2289
[9]   PLASTIC PH ELECTRODES FOR THE MEASUREMENT OF GASTROINTESTINAL PH [J].
RAWLINGS, JM ;
LUCAS, ML .
GUT, 1985, 26 (02) :203-207
[10]  
RAWLINGS JM, 1987, J GASTROENTEROLOGY, V22, P337