COMPARISON OF FILM AND ION-CHAMBER SYSTEMS FOR DEPTH-DOSE MEASUREMENTS FOR A 25-MV BEAM

被引:8
作者
ANDERSON, DW
STGEORGE, F
机构
[1] Department of Radiological Sciences, University of Oklahoma Health Sciences Center, Oklahoma City
关键词
D O I
10.1088/0031-9155/24/3/015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
It is shown that the use of film with a unit density rubber phantom can produce central axis percentage depth-dose values as much as 5% larger than ion chamber/water phantom values even for a 25 MV photon beam. It is suggested that care should be taken in interpreting film generated depth-dose data for photon beams even at the highest clinical energies.