COMPARISON OF NAL(TL), CDTE, AND HGL2 SURGICAL PROBES - PHYSICAL CHARACTERIZATION

被引:19
作者
BARBER, HB
BARRETT, HH
HICKERNELL, TS
KWO, DP
WOOLFENDEN, JM
ENTINE, G
BACCASH, CO
机构
[1] UNIV ARIZONA,CTR OPT SCI,TUCSON,AZ 85721
[2] RADIAT MONITORING DEVICES INC,WATERTOWN,MA 02172
[3] EG&G ENERGY MEASUREMENTS INC,GOLETA,CA 93017
关键词
D O I
10.1118/1.596717
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The physical properties of three surgical probes containing different radiation 'detectors are compared: a NaI(Tl) scintillator with a flexible, fiber-optic light guide, and two semiconductor detectors that operate at room temperature, CdTe and HgI2. Also compared are spectra, energy resolutions, and counting efficiencies measured at a variety of gamma-ray energies between 30 and 1000 keV. The energy resolution of the NaI probe is substantially poorer than that of either semiconductor probe due in part to light losses in coupling the scintillator to the fiber optics. The semiconductor probes have complex spectral response due to charge-carrier trapping and K x-ray escape, and not all photoelectric interactions in these detectors contribute to the useful part of the photopeak. Above 120 keV the counting efficiency for the NaI probe is an order of magnitude higher than for the CdTe and HgI2 probes. Both energy resolution and counting efficiency are slightly better for the HgI2 probe than for the CdTe probe.
引用
收藏
页码:373 / 381
页数:9
相关论文
共 32 条
[1]   GAMMA RESPONSE OF SEMI-INSULATING MATERIAL IN PRESENCE OF TRAPPING AND DETRAPPING [J].
AKUTAGAWA, W ;
ZANIO, K .
JOURNAL OF APPLIED PHYSICS, 1969, 40 (09) :3838-+
[2]   COMPARISON OF INVIVO SCINTILLATION PROBES AND GAMMA CAMERAS FOR DETECTION OF SMALL, DEEP TUMORS [J].
BARBER, HB ;
BARRETT, HH ;
WOOLFENDEN, JM ;
MYERS, KJ ;
HICKERNELL, TS .
PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (06) :727-739
[3]  
Browne E., 1986, TABLE RADIOACTIVE IS
[4]   FABRICATION AND EVALUATION OF HGI2 MINIATURE PROBES FOR INVIVO MEDICAL APPLICATIONS [J].
CAINE, S ;
HOLZER, A ;
BEINGLASS, I ;
SCHIEBER, M ;
LOWENTHAL, E .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1978, 25 (01) :649-652
[5]  
EVANDENBERG L, 1978, IEEE T NUCL SCI, V25, P395
[6]  
GARCIA DA, 1974, J NUCL MED, V15, P892
[7]  
HARCKE H T, 1988, Journal of Nuclear Medicine, V29, P881
[8]  
Harris CC, 1956, NUCLEONICS, V14
[9]   RADIOTELEMETRY PILL FOR MEASUREMENT OF IONIZING-RADIATION USING A MERCURIC IODIDE DETECTOR [J].
HASSAN, MA ;
PEARCE, G ;
EDWARDS, JPN .
PHYSICS IN MEDICINE AND BIOLOGY, 1978, 23 (02) :302-308
[10]  
HICKERNELL TS, 1988, J NUCL MED, V29, P1101