CATALYTIC CATHODIC STRIPPING VOLTAMMETRY OF OXIDIZED GLUTATHIONE AT A HANGING MERCURY DROP ELECTRODE IN THE PRESENCE OF NICKEL ION

被引:47
作者
BANICA, FG [1 ]
FOGG, AG [1 ]
MOREIRA, JC [1 ]
机构
[1] LOUGHBOROUGH UNIV TECHNOL,DEPT CHEM,LOUGHBOROUGH LE11 3TU,LEICS,ENGLAND
关键词
D O I
10.1016/0039-9140(94)00232-H
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Oxidized glutathione (GSSG) can be determined after previous accumulation on the HMDE at E > -0.2 V (vs. the Ag/AgCl reference electrode). GSH is formed during the accumulation, possibly by a mercury-ion-assisted hydrolytic disproportionation of GSSG. In the subsequent cathodic scan GSH is released and catalyses the reduction of nickel ion, giving a peak located at -0.6 V. This enables the determination of GSSG by differential-pulse cathodic stripping voltammetry at pH 7.0 in the phosphate-acetate or MOPS buffer containing 0.5-1.0 mM Ni(II). The detection limit is 10 nM. The calibration graph is linear even in the presence of small amounts of human serum albumin, HSA. However, HSA increases the detection limit (20 nM for 3 x 10(-4)% HSA). Acetyl-cysteine in small excess or Cu(II) present as reagent impurity do not interfere. Glutathione, cysteine and similar compounds, which accumulate as mercury salts and form stable nickel complexes, will interfere. The method is put forward as a novel alternative stripping voltammetric method to those involving accumulation and determination as mercury or copper salts and complexes, in the knowledge that it may have advantages in particular analytical situations. In particular the method discriminates against compounds which accumulate as mercury salts but which do not form stable nickel complexes.
引用
收藏
页码:227 / 234
页数:8
相关论文
共 21 条
[1]   APPLICATION OF CATALYTIC STRIPPING VOLTAMMETRY FOR THE DETERMINATION OF ORGANIC SULFUR-COMPOUNDS AT A HANGING MERCURY DROP ELECTRODE - BEHAVIOR OF CYSTEINE, CYSTINE AND N-ACETYLCYSTEINE IN THE PRESENCE OF NICKEL ION [J].
BANICA, FG ;
MOREIRA, JC ;
FOGG, AG .
ANALYST, 1994, 119 (02) :309-318
[2]  
BANICA FG, IN PRESS ANALYST
[3]  
BREZINA M, 1958, POLAROGRAPHY MED BIO, P522
[4]  
DIACU E, 1992, REV ROUM CHIM, V37, P1389
[5]  
DIACU E, 1993, REV ROUM CHIM, V38, P1397
[6]   CATHODIC STRIPPING VOLTAMMETRY .1. DETERMINATION OF ORGANIC SULFUR-COMPOUNDS, FLAVINS AND PORPHYRINS AT THE SUB-MICROMOLAR LEVEL [J].
FLORENCE, TM .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1979, 97 (02) :219-236
[7]   CATHODIC STRIPPING VOLTAMMETRIC BEHAVIOR OF CYSTEINE AND CYSTINE IN THE PRESENCE OF CUPRIC IONS [J].
FORSMAN, U .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1981, 122 (MAY) :215-231
[8]  
FORSMAN U, 1984, ANAL CHIM ACTA, V166, P141
[9]   CATHODIC STRIPPING VOLTAMMETRY OF GLUTATHIONE IN NATURAL-WATERS [J].
LEGALL, AC ;
VAN DEN BERG, CMG .
ANALYST, 1993, 118 (11) :1411-1415
[10]  
Mader P, 1970, COLLECT CZECH CHEM C, V35, P552