UTILIZATION OF METALLIC PLATFORMS IN ELECTROTHERMAL VAPORIZATION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY

被引:21
作者
MARAWI, I [1 ]
OLSON, LK [1 ]
WANG, JS [1 ]
CARUSO, JA [1 ]
机构
[1] UNIV CINCINNATI,DEPT CHEM,CINCINNATI,OH 45221
关键词
HYDRIDE GENERATION; ELECTROTHERMAL VAPORIZATION; INDUCTIVELY COUPLED PLASMA MASS SPECTROMETRY; HYDRIDE TRAPPING; METALLIC PLATFORMS;
D O I
10.1039/ja9951000007
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method to improve the analytical performance of hydride trapping on Pd inside a graphite furnace and subsequent determination by ICP-MS was investigated, Strips of tungsten, tantalum, molybdenum and rhenium were coated (electroplated/sputtered)with palladium and used as platforms inside the graphite furnace. A typical pyrolytic graphite-coated graphite platform was also coated with Pd in the same manner, for performance comparisons, Scanning electron microscopy (SEM) and X-ray fluorescence spectrometry were used to characterize the surfaces, SEM showed a smoother Pd layer covering the metallic substrates when compared with the graphite, An analyte signal reproducibility study revealed that Ta, Re and Mo were not ideal substrates for this purpose, The As peaks obtained from trapping on the graphite platform were broadened with a decrease in intensity after a few firings, The Pd-sputtered tungsten platform showed superior performance over the other platforms, The linear dynamic range obtained using this platform was improved by one order of magnitude in the upper limit (from 1 to 100 ng ml(-1)) over that reported using a graphite platform in previous work. The Limit of detection of arsenic was found to be 0.01 ng ml(-1), within the same range obtained from the graphite platform.
引用
收藏
页码:7 / 14
页数:8
相关论文
共 28 条
[1]  
ALVEREZ GH, 1991, ANAL LETT, V24, P1695
[2]  
[Anonymous], 1986, BINARY ALLOY PHASE D
[3]   DIFFERENT SAMPLE INTRODUCTION SYSTEMS FOR THE SIMULTANEOUS DETERMINATION OF AS, SB AND SE BY MICROWAVE-INDUCED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BULSKA, E ;
TSCHOPEL, P ;
BROEKAERT, JAC ;
TOLG, G .
ANALYTICA CHIMICA ACTA, 1993, 271 (01) :171-181
[4]   DETERMINATION OF LEAD IN WINES BY HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY [J].
CACHO, J ;
FERREIRA, V ;
NERIN, C .
ANALYST, 1992, 117 (01) :31-33
[5]  
DOIDGE PS, 1989, J ANAL ATOM SPECTROM, V4, P251, DOI 10.1039/ja9890400251
[6]   HPLC COUPLED WITH IN-LINE PHOTOLYSIS, HYDRIDE GENERATION AND FLAME ATOMIC-ABSORPTION SPECTROMETRY FOR THE SPECIATION OF TIN IN NATURAL-WATERS [J].
EBDON, L ;
HILL, SJ ;
JONES, P .
TALANTA, 1991, 38 (06) :607-611
[8]  
HEITKEMPER DT, 1992, INDUCTIVELY COUPLED, P781
[9]   COMPARISON OF PNEUMATIC NEBULIZATION AND HYDRIDE GENERATION INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETRY FOR ISOTOPIC ANALYSIS OF SELENIUM [J].
JANGHORBANI, M ;
TING, BTG .
ANALYTICAL CHEMISTRY, 1989, 61 (07) :701-708
[10]   DECOMPOSITION OF ORGANOARSENIC COMPOUNDS BY USING A MICROWAVE-OVEN AND SUBSEQUENT DETERMINATION BY FLOW-INJECTION HYDRIDE GENERATION ATOMIC-ABSORPTION SPECTROMETRY [J].
LE, XC ;
CULLEN, WR ;
REIMER, KJ .
APPLIED ORGANOMETALLIC CHEMISTRY, 1992, 6 (02) :161-171