REPRODUCIBILITY OF NORMAL CORNEAL POWER MEASUREMENTS WITH A KERATOMETER, PHOTOKERATOSCOPE, AND VIDEO IMAGING-SYSTEM

被引:55
作者
HANNUSH, SB
CRAWFORD, SL
WARING, GO
GEMMILL, MC
LYNN, MJ
NIZAM, A
机构
[1] EMORY UNIV,DEPT OPHTHALMOL,1327 CLIFTON RD NE,ATLANTA,GA 30322
[2] EMORY UNIV,DEPT EPIDEMIOL & BIOSTAT,ATLANTA,GA 30322
关键词
D O I
10.1001/archopht.1990.01070060087055
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
To determine the reproducibility of currently available instruments for measuring corneal power, two investigators measured comparable locations on 18 normal human corneas using a keratometer (Bausch & Lomb, Rochester, NY), the Corneascope (Kera Corp, Santa Clara, Calif), and the Corneal Modeling System (Computed Anatomy Inc, New York, NY). (For the two keratoscopes, average powers around rings at comparable locations were used.) Comparisons made between instruments at the significance level of .05 indicated that the keratometer was more reproducible than the Corneal Modeling System and the Corneascope, and the Corneal Modeling System was more reproducible than the Corneascope. Of the 31 rings projected on the surface of each cornea by the Corneal Modeling System, rings 2 through 13 were read reasonably reproducibly in that 76% of the measurements on these rings differed by no more than 0.5 diopter. The Corneal Modeling System (software version 1.16) is 83% as reproducible as a keratometer reading at approximately the edge of the 3-mm central zone on normal human corneas and provides information about corneal topography in a more reproducible and visually useful manner than the other two instruments. © 1990, American Medical Association. All rights reserved.
引用
收藏
页码:539 / 544
页数:6
相关论文
共 21 条
[1]  
COHEN KL, 1984, INVEST OPHTH VIS SCI, V25, P323
[2]  
DABEZIES OH, 1984, CONTACT LENSES
[3]  
DINGELDEIN SA, 1988, CORNEA, V7, P170
[4]  
DOSS JD, 1981, ARCH OPHTHALMOL-CHIC, V99, P1261
[5]  
GORMLEY DJ, 1988, CORNEA, V7, P30
[6]   ACCURACY AND PRECISION OF KERATOMETRY, PHOTOKERATOSCOPY, AND CORNEAL MODELING ON CALIBRATED STEEL BALLS [J].
HANNUSH, SB ;
CRAWFORD, SL ;
WARING, GO ;
GEMMILL, MC ;
LYNN, MJ ;
NIZAM, A .
ARCHIVES OF OPHTHALMOLOGY, 1989, 107 (08) :1235-1239
[7]  
HARRIS DJ, 1989, OPHTHALMOLOGY, V96, P1597
[8]  
KLYCE SD, 1984, INVEST OPHTH VIS SCI, V25, P1426
[9]  
KLYCE SD, 1986, CORNEAL SURGERY THEO, P563
[10]   CORNEAL TOPOGRAPHY OF EARLY KERATOCONUS [J].
MAGUIRE, LJ ;
BOURNE, WM .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1989, 108 (02) :107-112