CHARACTERISTICS OF AN IMPROVED MAGNETIC-IMPLANT GUIDANCE-SYSTEM

被引:28
作者
MCNEIL, RG
RITTER, RC
WANG, B
LAWSON, MA
GILLIES, GT
WIKA, KG
QUATE, EG
HOWARD, MA
GRADY, MS
机构
[1] UNIV VIRGINIA,DEPT PHYS,CHARLOTTESVILLE,VA 22901
[2] WANG NMR INC,LIVERMORE,CA 94550
[3] STEROTAXIS INC,LIVERMORE,CA 94550
[4] UNIV VIRGINIA,DEPT BIOMED ENGN,CHARLOTTESVILLE,VA 22908
[5] UNIV VIRGINIA,DEPT MECH AEROSP & NUCL ENGN,CHARLOTTESVILLE,VA 22901
[6] UNIV VIRGINIA,DEPT COMP SCI,CHARLOTTESVILLE,VA 22901
[7] UNIV VIRGINIA,DEPT NUCL ENGN & ENGN PHYS,CHARLOTTESVILLE,VA 22901
[8] UNIV WASHINGTON,HARBORVIEW MED CTR,DEPT NEUROSURG,SEATTLE,WA 98104
关键词
D O I
10.1109/10.398641
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The previous companion paper described the motivation, design, and early experiments of a Magnetic Stereotaxis System. The part of the system considered in these papers is a helmet with a roughly cubic array of six superconducting coils used to apply force on small permanent magnet pellets in brain and in brain phantom material. This apparatus will be used to deliver drugs and other therapies directly into deep brain tissues, under control of a computer and fluoroscopic imaging system. Here, we analyze the general stability problems of controlling the currents in the coils for impulsive stepwise motion of the pellet, subject to quench avoidance in the superconducting coils, and in the face of Earnshaw's theorem governing stability in static magnetic fields. We also describe solutions that have been found to the primary difficulties limiting controlled pellet motion in the studies presented in the companion paper.
引用
收藏
页码:802 / 808
页数:7
相关论文
共 12 条
[1]  
BOZORTH RM, 1951, FERROMAGNETISM, P54
[2]   DEMAGNETIZING FACTORS FOR CYLINDERS [J].
CHEN, DX ;
BRUG, JA ;
GOLDFARB, RB .
IEEE TRANSACTIONS ON MAGNETICS, 1991, 27 (04) :3601-3619
[3]  
CHEUNG WS, 1983, 1983 P INT SCH S PRE, P81
[4]  
Earnshaw S., 1842, T CAMBRIDGE PHILOS S, V7, P97
[5]  
FISEL CR, 1986, THESIS STANFORD U
[6]  
LAWSON MA, 1991, P SOC PHOTO-OPT INS, V1445, P265, DOI 10.1117/12.45224
[7]   FUNCTIONAL DESIGN-FEATURES AND INITIAL PERFORMANCE-CHARACTERISTICS OF A MAGNETIC-IMPLANT GUIDANCE-SYSTEM FOR STEREOTAXIC NEUROSURGERY [J].
MCNEIL, RG ;
RITTER, RC ;
WANG, B ;
LAWSON, MA ;
GILLIES, GT ;
WIKA, KG ;
QUATE, EG ;
HOWARD, MA ;
GRADY, MS .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (08) :793-801
[8]   EXPERIMENTAL-DETERMINATION OF THE FORCE REQUIRED FOR INSERTION OF A THERMOSEED INTO DEEP BRAIN-TISSUES [J].
MOLLOY, JA ;
RITTER, RC ;
GRADY, MS ;
HOWARD, MA ;
QUATE, EG ;
GILLIES, GT .
ANNALS OF BIOMEDICAL ENGINEERING, 1990, 18 (03) :299-313
[9]  
RAMOS PA, 1991, P SOC PHOTO-OPT INS, V1443, P160, DOI 10.1117/12.43438
[10]   ELECTROOPTIC IMAGING CHAIN FOR A BIPLANAR FLUOROSCOPE FOR NEUROSURGERY - MAGNETIC-FIELD SENSITIVITY AND CONTRAST MEASUREMENTS [J].
RAMOS, PA ;
ALLISON, SW ;
MOLLOY, JA ;
LAWSON, MA ;
QUATE, EG ;
RITTER, RC ;
GILLIES, GT ;
GRADY, MS ;
HOWARD, MA .
OPTICAL ENGINEERING, 1993, 32 (07) :1644-1656