BLOOD - ARTIFICIAL SURFACE INTERACTIONS DURING CARDIOPULMONARY BYPASS - A COMPARATIVE-STUDY OF 4 OXYGENATORS

被引:12
作者
BENEDETTI, M
DECATERINA, R
BIONDA, A
GARDINALI, M
CICARDI, M
MAFFEI, S
GAZZETTI, P
PISTOLESI, P
VERNAZZA, F
MICHELASSI, C
GIORDANI, R
SALVATORE, L
机构
[1] CNR,PHYSIOL CLIN,DEPT CARDIAC SURG,ROME,ITALY
[2] UNIV PISA,MED CLIN 2,I-56100 PISA,ITALY
[3] UNIV MILAN,MED CLIN 5,I-20122 MILAN,ITALY
关键词
D O I
10.1177/039139889001300808
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Evaluation of the biocompatibility of four different types of oxygenator (bubble, membrane, hollow fibre and 'hybrid') was performed on 26 patients undergoing cardiopulmonary bypass during elective coronary surgery. More platelet derangement and an increased degree of hemolysis, revealed by higher plasmatic concentration of beta-thromboglobulin, platelet factor 4 and plasmatic free hemoglobin (p<0.05), was seen when using the bubble oxygenator. Damage to blood cells was minimal with the membrane oxygenator while the 'hybrid' and the hollow fibre oxygenators proved to rank at an intermediate level. Complement activation at the beginning of the cardiopulmonary bypass occured via the alternative pathway as demonstrated by C3a [(des arg]) increase (up to nine times) without a concomitant elevation of C4a[(des arg.]) Cardiopulmonary bypass complement activation was quantitatively similar with all the oxygenators. A further activation via the classical pathway occurred in all the patients after protamine injection. Consistent differences as far as clinical and biological effects exist among the various commercially available cardiopulmonary bypass apparatus; our study provides guidelines for the evaluation and selection of devices which might reduce postoperative sequelae.
引用
收藏
页码:488 / 497
页数:10
相关论文
共 46 条
  • [1] ADDONIZIO VP, 1979, J THORAC CARDIOV SUR, V77, P119
  • [2] [Anonymous], 1971, STAT PRINCIPLES EXPT
  • [3] ARCIDI JM, 1988, J THORAC CARDIOV SUR, V95, P773
  • [4] BENEDETTI M, 1988, CAH CECEC, V30, P303
  • [5] BIONDA A, 1986, IMMUNE METABOLIC ASP, P246
  • [6] 2 MEMBRANE OXYGENATORS AND A BUBBLER - A CLINICAL COMPARISON
    BOERS, M
    VANDENDUNGEN, JJAM
    KARLICZEK, GF
    BRENKEN, U
    VANDERHEIDE, JNH
    WILDEVUUR, CRH
    [J]. ANNALS OF THORACIC SURGERY, 1983, 35 (04) : 455 - 462
  • [7] PERIOPERATIVE CHANGES IN COMPLEMENT ASSOCIATED WITH CARDIOPULMONARY BYPASS
    BORALESSA, H
    SHIFFERLI, JA
    ZAIMI, F
    WATTS, E
    WHITWAM, JG
    REES, AJ
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1982, 54 (10) : 1047 - 1052
  • [8] CAVAROCCHI NC, 1986, CIRCULATION, V74, P130
  • [9] CAVAROCCHI NC, 1985, SURGERY, V98, P525
  • [10] COMPLEMENT ACTIVATION DURING CARDIOPULMONARY BYPASS - EVIDENCE FOR GENERATION OF C3A AND C5A ANAPHYLATOXINS
    CHENOWETH, DE
    COOPER, SW
    HUGLI, TE
    STEWART, RW
    BLACKSTONE, EH
    KIRKLIN, JW
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1981, 304 (09) : 497 - 503