CLINICAL-EXPERIENCE WITH A MULTIELEMENT ULTRASONIC HYPERTHERMIA SYSTEM - ANALYSIS OF TREATMENT TEMPERATURES

被引:46
作者
SAMULSKI, TV
GRANT, WJ
OLESON, JR
LEOPOLD, KA
DEWHIRST, MW
VALLARIO, P
BLIVIN, J
机构
[1] Duke University Medical Center, Radiation Oncology, Durham, NC 27710
关键词
D O I
10.3109/02656739009140972
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A summary of tumour temperature data obtained from 31 patients who underwent 147 hyperthermia treatments with the Sonotherm 1000 ultrasonic system is presented. The treatment goal was to achieve a minimum of 42.0°C in tumour for 60 min duration with normal tissues remaining below 43.0°C. In 83% of treatments at least one measured tumour temperature reached or exceeded 42.0°C at some time during the treatment. Nineteen per cent of these treatments had a time-and spatial-averaged temperature (measured in tumour) ≥ 42.0°C. A variety of anatomical sites were treated and these were grouped into four categories: groin/trunk, axilla, breast/chest wall and head/neck. Measured temperatures in tumours located in the groin and trunk sites were significantly higher (22% ≥ 42°C) than other locations. The head and neck treatment temperatures were significantly lower (8% of measured points ≥ 42°C). © 1990 Informa UK Ltd All rights reserved: reproduction in whole or part not permitted.
引用
收藏
页码:909 / 922
页数:14
相关论文
共 18 条
[1]   ULTRASONIC HYPERTHERMIA AND RADIATION IN THE MANAGEMENT OF INTRAOCULAR MALIGNANT-MELANOMA [J].
COLEMAN, DJ ;
LIZZI, FL ;
BURGESS, SEP ;
SILVERMAN, RH ;
SMITH, ME ;
DRILLER, J ;
ROSADO, A ;
ELLSWORTH, RM ;
HAIK, BG ;
ABRAMSON, DH ;
MCCORMICK, B .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 1986, 101 (06) :635-642
[2]   HUMAN CANCER-TREATMENT WITH ULTRASOUND [J].
CORRY, PM ;
JABBOURY, K ;
ARMOUR, EP ;
KONG, JS .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1984, 31 (05) :444-456
[3]   ULTRASOUND-INDUCED HYPERTHERMIA FOR THE TREATMENT OF HUMAN SUPERFICIAL TUMORS [J].
CORRY, PM ;
BARLOGIE, B ;
TILCHEN, EJ ;
ARMOUR, EP .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1982, 8 (07) :1225-1229
[4]   EVALUATION OF EQUIPMENT FOR HYPERTHERMIC TREATMENT OF CANCER [J].
CORRY, PM ;
JABBOURY, K ;
KONG, JS ;
ARMOUR, EP ;
MCCRAW, FJ ;
LEDUC, T .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (01) :53-74
[5]   AUTOMATED TEMPERATURE SCANNING FOR HYPERTHERMIA TREATMENT MONITORING [J].
ENGLER, MJ ;
DEWHIRST, MW ;
WINGET, JM ;
OLESON, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1987, 13 (09) :1377-1382
[6]  
FESSENDEN P, 1984, CANCER RES, V44, P4799
[7]  
HAND J, 1986, PHYSICAL TECHNIQUES
[8]  
HYNYNEN K, 1987, 9TH P IEEE ANN C ENG, V3, P1644
[9]   STANFORD-UNIVERSITY INSTITUTIONAL REPORT .1. EVALUATION OF EQUIPMENT FOR HYPERTHERMIA TREATMENT OF CANCER [J].
KAPP, DS ;
FESSENDEN, P ;
SAMULSKI, TV ;
BAGSHAW, MA ;
COX, RS ;
LEE, ER ;
LOHRBACH, AW ;
MEYER, JL ;
PRIONAS, SD .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (01) :75-115
[10]  
LELE PP, 1983, HYPERTHERMIA CANCER, P333