NUCLEAR-MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A PACEMAKER

被引:31
作者
INBAR, S [1 ]
LARSON, J [1 ]
BURT, T [1 ]
MAFEE, M [1 ]
EZRI, MD [1 ]
机构
[1] UNIV ILLINOIS,DEPT RADIOL,CHICAGO,IL 60680
关键词
MRI; PACEMAKER; AV BLOCK; RADIOFREQUENCY INTERFERENCE;
D O I
10.1097/00000441-199303000-00009
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The presence of a pacemaker has been considered a contraindication for magnetic resonance imaging (MRI) examination. The authors describe a patient with a pacemaker who underwent MRI of the head without sequelae. Spine-echo MRI was preformed with 1.5 Tesla imager using head coil only. Prior to imaging, the pacemaker was programmed to the OOO mode. In carefully selected pacing-independent patients, with a pacemaker in a bipolar configuration, MRI of the head appears to be safe provided the pacemaker is programmed to 000 and enough scar tissue has developed to prevent pacemaker or lead movement.
引用
收藏
页码:174 / 175
页数:2
相关论文
共 13 条
[1]   NUCLEAR MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A DDD PACEMAKER [J].
ALAGONA, P ;
TOOLE, JC ;
MANISCALCO, BS ;
GLOVER, MU ;
ABERNATHY, GT ;
PRIDA, XE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04) :619-619
[2]  
BRDLEY WG, 1986, MAGN RESON ANNU, P81
[3]  
DELUCA SA, 1990, AM FAM PHYSICIAN, V41, P145
[4]   EFFECT OF MAGNETIC-RESONANCE-IMAGING ON DDD PACEMAKERS [J].
ERLEBACHER, JA ;
CAHILL, PT ;
PANNIZZO, F ;
KNOWLES, RJR .
AMERICAN JOURNAL OF CARDIOLOGY, 1986, 57 (06) :437-440
[5]   THE EFFECTS OF NUCLEAR MAGNETIC-RESONANCE IMAGERS ON EXTERNAL AND IMPLANTABLE PULSE GENERATORS [J].
FETTER, J ;
ARAM, G ;
HOLMES, DR ;
GRAY, JE ;
HAYES, DL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1984, 7 (04) :720-727
[6]  
FURMAN S, 1989, PRACTICE PACING, P606
[7]   EFFECT OF 1.5 TESLA NUCLEAR-MAGNETIC-RESONANCE IMAGING SCANNER ON IMPLANTED PERMANENT PACEMAKERS [J].
HAYES, DL ;
HOLMES, DR ;
GRAY, JE .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (04) :782-786
[8]   THE EFFECTS OF MAGNETIC-RESONANCE-IMAGING ON IMPLANTABLE PULSE GENERATORS [J].
HOLMES, DR ;
HAYES, DL ;
GRAY, JE ;
MERIDETH, J .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (03) :360-370
[9]  
NEUFANG KFR, 1989, FORTSCHR RONTG NEUEN, V151, P659
[10]   THE EFFECTS OF NUCLEAR MAGNETIC-RESONANCE ON PATIENTS WITH CARDIAC-PACEMAKERS [J].
PAVLICEK, W ;
GEISINGER, M ;
CASTLE, L ;
BORKOWSKI, GP ;
MEANEY, TF ;
BREAM, BL ;
GALLAGHER, JH .
RADIOLOGY, 1983, 147 (01) :149-153