AN AUTOMATED SEIZURE MONITORING-SYSTEM FOR PATIENTS WITH INDWELLING RECORDING ELECTRODES

被引:34
作者
HARDING, GW [1 ]
机构
[1] WASHINGTON UNIV, SCH MED, MCDONNELL CTR STUDIES HIGHER BRAIN FUNCT, ST LOUIS, MO 63110 USA
来源
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY | 1993年 / 86卷 / 06期
关键词
SEIZURE MONITOR; AUTOMATIC SEIZURE DETECTION; INDWELLING ELECTRODES;
D O I
10.1016/0013-4694(93)90138-L
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
An automated monitoring system has been developed to record from indwelling electrode arrays in patients undergoing evaluation for surgical treatment of intractable seizures. The functional aspects of this system's design are discussed and the range of electrical seizure patterns, other epileptiform events and artifacts that the system must handle are described. The system includes a flowing graphics image of as many as 32 channels of real-time ECoG, automatic seizure detection, recording of the ECoG for up to 6 min prior to seizure onset, and EEG machine start and stop control. In 40 monitored patients, the automated system achieved a seizure detection and recording accuracy of 95% for 792 clinical and subclinical seizures during 1578 h of monitoring with an artifact rate of 28% for all events recorded. However, system performance was judged upon the 86% accuracy for detection and recording of patient-specific seizure patterns (multiple seizures with the same pattern counted as 1) with 1.26 brief spike bursts and 0.67 artifacts/h.
引用
收藏
页码:428 / 437
页数:10
相关论文
共 17 条
[1]   ELECTRONIC CIRCUIT FOR DETECTION OF EEG SEIZURES RECORDED WITH IMPLANTED ELECTRODES [J].
BABB, TL ;
MARIANI, E ;
CRANDALL, PH .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1974, 37 (03) :305-308
[2]   CONTEXT-BASED AUTOMATED DETECTION OF EPILEPTOGENIC SHARP TRANSIENTS IN THE EEG - ELIMINATION OF FALSE POSITIVES [J].
GLOVER, JR ;
RAGHAVAN, N ;
KTONAS, PY ;
FROST, JD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1989, 36 (05) :519-527
[4]   SURGICAL-MANAGEMENT OF EPILEPSY USING EPIDURAL RECORDINGS TO LOCALIZE THE SEIZURE FOCUS - REVIEW OF 100 CASES [J].
GOLDRING, S ;
GREGORIE, EM .
JOURNAL OF NEUROSURGERY, 1984, 60 (03) :457-466
[5]  
Goldring S, 1987, SURGICAL TREATMENT E, P445
[6]   AUTOMATIC RECOGNITION AND QUANTIFICATION OF INTERICTAL EPILEPTIC ACTIVITY IN HUMAN SCALP EEG [J].
GOTMAN, J ;
GLOOR, P .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1976, 41 (05) :513-529
[7]  
Gotman J, 1985, Electroencephalogr Clin Neurophysiol Suppl, V37, P133
[8]   AUTOMATIC RECOGNITION OF EPILEPTIC SEIZURES IN THE EEG [J].
GOTMAN, J .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1982, 54 (05) :530-540
[9]   AUTOMATIC SEIZURE DETECTION - IMPROVEMENTS AND EVALUATION [J].
GOTMAN, J .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1990, 76 (04) :317-324
[10]   AUTOMATED ONLINE, REAL-TIME LABORATORY FOR SINGLE NEURON STUDIES [J].
HARDING, GW ;
TOWE, AL .
COMPUTERS AND BIOMEDICAL RESEARCH, 1976, 9 (05) :471-501